Over 15 years experience of providing support of MEAN WELL Power Supplies

Image shown is a representation only. Exact specifications should be obtained from the product datasheet.

×
High Resolution Image

Mean Well RPS-500-27 Green Power Supply W 27V 11.9A - Medical Power Supply

5" x 3" compact size, medical safety approved, extreme low leakage current

|

Description

Features

5"x3" compact size
320W convection, 500W force air
550W peak power (3 sec.)/td>
Medical safety approved (2xMOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
MEMI for both Class I & Class II configuration
-30~+70°C wide range operating temperature
No load power consumption under 0.5W by PS-ON control
High efficiency up to 94%
Protections: Short circuit / Overload / Over voltage / Over Temp.
5Vdc standby output, 12Vdc fan supply, Power Good, Power Fail, and remote sense
Operating altitude up to 4000 meters (Note.5)
LED indicator for power on
3 years warranty

General Specification 

AC input voltage range 80 ~ 264VAC | 113 ~ 370VDC
AC inrush current COLD START 40A/115VAC 80A/230VAC
DC adjustment range ±10% rated output voltage
Overload protection 105%~135% hiccup mode, auto-recovery
Over-voltage protection CH1: 115%~135% shut off, re-power on to recovery
Withstand voltage I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
Working temperature -30~+70°C (refer to output derating curve)
Leakage current Earth leakage current <220 A/264VAC 50Hz, Touch current < 100 A/264VAC
Safety standards ANSI/AAMI ES60601-1, TUV EN60601-1 approved
EMC standards EN55011/EN55022 class B, EN61000-4-2,3,4,5,6,8,11, EN61000-3-2,3, EN60601-1-2, EN61000-6-2 (RPS(G)-160)
Connection 3+8+2(4)P/3.96mm pitch JST B3P / B8P-VH, B2B (B4B)-XH
Setup, rise, hold up time 1000ms, 30ms, 10ms at full load and 230VAC
Dimension (LxWxH) 127 x 76.2 x 41mm | 5 x 3 x 1.61 inch

Specifications

  • Brand
    Mean Well
  • Series
    RPS Series
  • Power Rating
    321.3 Watt
  • Output Voltage
    27 Volt
  • Rated Current
    11.9 Ampere
  • Input Voltage
    80V~264V
  • Standards Met
    CB,CE,cUL,EAC,SHIDE
  • Operation Mode
    Constant Voltage
  • Operating Temperature
    -30°C~70°C
  • Warranty
    3 Year
  • Length
    12.7 cm
  • Width
    7.62 cm
  • Height
    4.1 cm
  • Weight
    0.46 kg

File Downloads

::specification::

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8431134900523, "variants" :[{ "context":"default", "variant_id": 45665393049899, "sku":"RPS-500-27", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_identity\/redirect?locale=en\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well RPS-500-27 Green Power Supply W 27V 11.9A - Medical Power Supply
Mean Well

Mean Well RPS-500-27 Green Power Supply W 27V 11.9A - Medical Power Supply

Features

5"x3" compact size
320W convection, 500W force air
550W peak power (3 sec.)/td>
Medical safety approved (2xMOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1
MEMI for both Class I & Class II configuration
-30~+70°C wide range operating temperature
No load power consumption under 0.5W by PS-ON control
High efficiency up to 94%
Protections: Short circuit / Overload / Over voltage / Over Temp.
5Vdc standby output, 12Vdc fan supply, Power Good, Power Fail, and remote sense
Operating altitude up to 4000 meters (Note.5)
LED indicator for power on
3 years warranty

General Specification 

AC input voltage range 80 ~ 264VAC | 113 ~ 370VDC
AC inrush current COLD START 40A/115VAC 80A/230VAC
DC adjustment range ±10% rated output voltage
Overload protection 105%~135% hiccup mode, auto-recovery
Over-voltage protection CH1: 115%~135% shut off, re-power on to recovery
Withstand voltage I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
Working temperature -30~+70°C (refer to output derating curve)
Leakage current Earth leakage current <220 A/264VAC 50Hz, Touch current < 100 A/264VAC
Safety standards ANSI/AAMI ES60601-1, TUV EN60601-1 approved
EMC standards EN55011/EN55022 class B, EN61000-4-2,3,4,5,6,8,11, EN61000-3-2,3, EN60601-1-2, EN61000-6-2 (RPS(G)-160)
Connection 3+8+2(4)P/3.96mm pitch JST B3P / B8P-VH, B2B (B4B)-XH
Setup, rise, hold up time 1000ms, 30ms, 10ms at full load and 230VAC
Dimension (LxWxH) 127 x 76.2 x 41mm | 5 x 3 x 1.61 inch