RPS-75-12 is the 12V output model of this series: rated current 6.3A, rated power 75.6W, typical efficiency 82%.
RPS-75 is a 75W highly reliable green PCB type medical power supply with a high power density on the 5” by 3” footprint. It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 86% and the extremely low no load power consumption is down below 0.75W. RPS-75 is able to be used for ClassⅠ (with FG) system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.
Features
- 5"×3" compact size
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- 75W convection,100W force air
- EMI Class B for ClassⅠconfiguration
- No load power consumption<0.75W
- Remote sense functiom
- Protections: Short circuit / Overload / Over voltage
- Lifetime > 80K hours
- Operating altitude up to 3000 meters
- 3 years warranty
Specifications
OUTPUT
- DC VOLTAGE
- 12V
- RATED CURRENT
- 6.3A
- CURRENT RANGE
- 0 ~ 8.3A
- RATED POWER
- 75.6W
- PEAK LOAD (23.5CFM)
- 99.6W
- RIPPLE & NOISE (max.)Note.2
- 60mVp-p
- VOLTAGE ADJ. RANGE
- 11.4 ~ 13.2V
- VOLTAGE TOLERANCENote.3
- ±1.0%
- LINE REGULATION
- ±0.5%
- LOAD REGULATION
- ±1.0%
- SETUP, RISE TIME
- 500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
- HOLD UP TIME (Typ.)
- 90ms/230VAC 20ms/115VAC at full load
INPUT
- VOLTAGE RANGE
- 90 ~ 264VAC 127 ~370VDC
- FREQUENCY RANGE
- 47 ~ 63Hz
- EFFICIENCY(Typ.)
- 82%
- AC CURRENT (Typ.)
- 1.5A/115VAC 1A/230VAC
- INRUSH CURRENT (Typ.)
- COLD START 25A/115VAC 50A/230VAC
- LEAKAGE CURRENT(max.)Note.4
- Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC
PROTECTION
- OVERLOAD
- 140 ~ 180% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed.
- OVER VOLTAGE
- 13.8 ~ 16.2V Protection type : Shut down o/p voltage, re-power to recover
ENVIRONMENT
- WORKING TEMP.
- -20 ~ +70℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20 ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -40 ~ +85℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- ±0.03%/℃ (0 ~ 50℃)
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
- OPERATING ALTITUDENote.5
- 3000 meters
SAFETY & EMC
- SAFETY STANDARDS
- IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2 CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)
- ISOLATION LEVEL
- Primary-Secondary:2xMOPP, Primary-Earth:1xMOPP
- WITHSTAND VOLTAGE
- I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
- ISOLATION RESISTANCE
- I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
- EMC EMISSION
-
Parameter Standard Test Level Conducted emission BS EN/EN55011 (CISPR11) Class B Radiated emission BS EN/EN55011 (CISPR11) Class B Harmonic current BS EN/EN61000-3-2 Class A Voltage flicker BS EN/EN61000-3-3 - EMC IMMUNITY
-
BS EN/EN55035, BS EN/EN60601-1-2
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 4, 15KV air; Level 4, 8KV contact RF field susceptibility BS EN/EN61000-4-3 Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz ) EFT bursts BS EN/EN61000-4-4 Level 3, 2KV Surge susceptibility BS EN/EN61000-4-5 Level 4, 4KV/Line-FG; 2KV/Line-Line Conducted susceptibility BS EN/EN61000-4-6 Level 3, 10V Magnetic field immunity BS EN/EN61000-4-8 Level 4, 30A/m Voltage dip, interruption BS EN/EN61000-4-11 100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods
OTHERS
- MTBF
- 2837.6K hrs min. Telcordia SR-332 (Bellcore) ; 446.8K hrs min. MIL-HDBK-217F (25℃)
- DIMENSION (L*W*H)
- 127*76.2*31mm or 5" * 3" *1.22" inch
- PACKING
- 0.26Kg; 63pcs/16.3Kg/1.28CUFT
Test conditions and notes7
- 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
- 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
- 3. Tolerance : includes set up tolerance, line regulation and load regulation.
- 4. Touch current was measured from primary input to DC output.
- 5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
- 6. Heat Sink HS1,HS2,HS3 can not be shorted.
- 7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
Technical drawings
Typical applications
- Oral irrigator
- Hemodialysis machine
- Medical computer monitors
- Sleep apnea devices
Downloads
FAQ
Can I use the RPS-75-12 to power a 12V industrial control panel or HMI?
Yes, if the control panel or HMI requires a 12V constant voltage supply and stays within 6.3A and 75.6W. The wide 90-264VAC input also fits facilities that may use different AC mains ranges.
Is this 12V power supply suitable for embedded electronics or single-board computer systems?
It can be a fit for embedded or computing hardware that needs regulated 12V DC and draws no more than 6.3A. Because it is a constant voltage model rated at 75.6W, it is better matched to stable 12V loads than to devices needing another output voltage.
Can the RPS-75-12 be used in low-power medical or laboratory equipment?
It is intended for equipment that needs 12V output up to 6.3A and benefits from medical safety approval. Buyers should still confirm the end device's exact 12V load stays within the 75.6W rating.
How does the RPS-75-12 compare with the other RPS-75 output options?
Within the RPS-75 group, the output options are 3.3V 15A 49.5W, 5V 14A 70W, 12V 6.3A 75.6W, 15V 5A 75W, 24V 3.2A 76.8W, 36V 2.1A 75.6W, and 48V 1.6A 76.8W. The RPS-75-12 is the 12V member of that set, with mid-range current and 75.6W total output.
Is the RPS-75-12 suitable for outdoor or harsh-environment use?
Its operating temperature range is -20°C to +70°C, so the temperature tolerance is broad for many protected installations. However, no IP rating is listed in the provided specs, so it is better treated as an indoor component power supply for protected enclosures rather than exposed outdoor use.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other RPS-75 Models
6 other models| Model | Power | Voltage | Current | |
|---|---|---|---|---|
| Mean Well RPS-75-15 Green Power Supply 75W 15V 5A - Medical Power Supply | 75W | 15V | 5A | View |
| Mean Well RPS-75-5 Green Power Supply 70W 5V 14A - Medical Power Supply | 70W | 5V | 14A | View |
| Mean Well RPS-75-48 Green Power Supply 76.8W 48V 1.6A - Medical Power Supply | 76.8W | 48V | 1.6A | View |
| Mean Well RPS-75-36 Green Power Supply 75.6W 36V 2.1A - Medical Power Supply | 75.6W | 36V | 2.1A | View |
| Mean Well RPS-75-3.3 Green Power Supply 49.5W 3.3V 15A - Medical Power Supply | 49.5W | 3.3V | 15A | View |
| Mean Well RPS-75-24 Green Power Supply 76.8W 24V 3.2A - Medical Power Supply | 76.8W | 24V | 3.2A | View |
Please log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.