Mean Well RPS-75-36 Green Power Supply 75.6W 36V 2.1A - Medical Power Supply

5" x 3" compact sizemedical safety approvedno load power consumption < 0.75Wlifetime > 80K hours
SKU RPS-75-36
Log in to see price Trade pricing & online ordering for account holders
Available to order
|

Overview

RPS-75-36 is the 36V output model of this series: rated current 2.1A, rated power 75.6W, typical efficiency 86%.

RPS-75 is a 75W highly reliable green PCB type medical power supply with a high power density on the 5” by 3” footprint. It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 86% and the extremely low no load power consumption is down below 0.75W. RPS-75 is able to be used for ClassⅠ (with FG) system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Features

  • 5"×3" compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • 75W convection,100W force air
  • EMI Class B for ClassⅠconfiguration
  • No load power consumption<0.75W
  • Remote sense functiom
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 80K hours
  • Operating altitude up to 3000 meters
  • 3 years warranty

Specifications

RPS-75-36Specifications from the official datasheet (2024-11-22 edition)

OUTPUT

DC VOLTAGE
36V
RATED CURRENT
2.1A
CURRENT RANGE
0 ~ 2.8A
RATED POWER
75.6W
PEAK LOAD (23.5CFM)
100.8W
RIPPLE & NOISE (max.)Note.2
100mVp-p
VOLTAGE ADJ. RANGE
34.2 ~ 39.6V
VOLTAGE TOLERANCENote.3
±1.0%
LINE REGULATION
±0.5%
LOAD REGULATION
±1.0%
SETUP, RISE TIME
500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
90ms/230VAC 20ms/115VAC at full load

INPUT

VOLTAGE RANGE
90 ~ 264VAC 127 ~370VDC
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY(Typ.)
86%
AC CURRENT (Typ.)
1.5A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)
COLD START 25A/115VAC 50A/230VAC
LEAKAGE CURRENT(max.)Note.4
Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC

PROTECTION

OVERLOAD
140 ~ 180% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed.
OVER VOLTAGE
41.4 ~ 48.6V Protection type : Shut down o/p voltage, re-power to recover

ENVIRONMENT

WORKING TEMP.
-20 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
±0.03%/℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDENote.5
3000 meters

SAFETY & EMC

SAFETY STANDARDS
IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2 CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL
Primary-Secondary:2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
ParameterStandardTest Level
Conducted emissionBS EN/EN55011 (CISPR11)Class B
Radiated emissionBS EN/EN55011 (CISPR11)Class B
Harmonic currentBS EN/EN61000-3-2Class A
Voltage flickerBS EN/EN61000-3-3
EMC IMMUNITY

BS EN/EN55035, BS EN/EN60601-1-2

ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 4, 15KV air; Level 4, 8KV contact
RF field susceptibilityBS EN/EN61000-4-3Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT burstsBS EN/EN61000-4-4Level 3, 2KV
Surge susceptibilityBS EN/EN61000-4-5Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibilityBS EN/EN61000-4-6Level 3, 10V
Magnetic field immunityBS EN/EN61000-4-8Level 4, 30A/m
Voltage dip, interruptionBS EN/EN61000-4-11100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods

OTHERS

MTBF
2837.6K hrs min. Telcordia SR-332 (Bellcore) ; 446.8K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (L*W*H)
127*76.2*31mm or 5" * 3" *1.22" inch
PACKING
0.26Kg; 63pcs/16.3Kg/1.28CUFT
Test conditions and notes7
  1. 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
  3. 3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. 4. Touch current was measured from primary input to DC output.
  5. 5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  6. 6. Heat Sink HS1,HS2,HS3 can not be shorted.
  7. 7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

Technical drawings

RPS-75-36 Block Diagram
Block Diagram
RPS-75-36 Derating Curve
Derating Curve
RPS-75-36 Mechanical Specification
Mechanical Specification

Typical applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

Downloads

FAQ

Can I use the RPS-75-36 for 36V automation or industrial control equipment?

Yes, if the equipment requires 36V constant voltage and draws no more than 2.1A or 75.6W. It is better matched to lighter-current 36V control loads than to higher-current 36V systems.

Is the RPS-75-36 suitable for 36V test, measurement, or instrumentation devices?

It can be used for instrumentation or measurement equipment that specifies 36V DC input and stays within the 2.1A output limit. The wide 90-264VAC input range also supports installations across different AC supply standards.

Can this model power compact medical or laboratory equipment that needs 36V DC?

Yes, for devices that require a 36V constant voltage source and consume up to 75.6W. Its medical safety approval is relevant for medical or lab designs, but the load still has to remain within the 2.1A rating.

How does the RPS-75-36 differ from other RPS-75 output versions?

The RPS-75 lineup includes 3.3V 15A 49.5W, 5V 14A 70W, 12V 6.3A 75.6W, 15V 5A 75W, 24V 3.2A 76.8W, 36V 2.1A 75.6W, and 48V 1.6A 76.8W models. The RPS-75-36 is the 36V member and is rated for 2.1A output.

Is the RPS-75-36 appropriate for outdoor or washdown environments?

The unit is specified for operation from -20°C to +70°C. It is an open-frame PCB type power supply with no IP rating, so it must be installed in a protected indoor enclosure and is not suitable for exposed outdoor or washdown locations.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8431165473067, "variants" :[{ "context":"default", "variant_id": 45665593852203, "sku":"RPS-75-36", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well RPS-75-36 Green Power Supply W 36V 2.1A - Medical Power Supply
Mean Well

Mean Well RPS-75-36 Green Power Supply 75.6W 36V 2.1A - Medical Power Supply

RPS-75-36 is the 36V output model of this series: rated current 2.1A, rated power 75.6W, typical efficiency 86%.

RPS-75 is a 75W highly reliable green PCB type medical power supply with a high power density on the 5” by 3” footprint. It accepts 90~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 86% and the extremely low no load power consumption is down below 0.75W. RPS-75 is able to be used for ClassⅠ (with FG) system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Features

  • 5"×3" compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • 75W convection,100W force air
  • EMI Class B for ClassⅠconfiguration
  • No load power consumption<0.75W
  • Remote sense functiom
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 80K hours
  • Operating altitude up to 3000 meters
  • 3 years warranty

Typical applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices
View product