Mean Well PM-15-5 Power Supply 15W 5V 3A

SKU PM-15-5
Log in to see price Trade pricing & online ordering for account holders
Available to order
|

Overview

PM-15-5 is the 5V output model of this series: rated current 3A, rated power 15W, typical efficiency 76%.

Features

  • Universal AC input / Full range
  • Protections: Short circuit / Overload / Over voltage
  • Ultra-miniature size, light weight
  • Cooling by free air convection
  • Isolation classⅡ
  • Medical safety approved (2 x MOPP between primary to secondary)
  • No load power consumption<0.5W
  • 100% full load burn-in test
  • Fixed switching frequency at 67KHz
  • High reliability
  • Suitable for BF application with appropriate system consideration
  • 3 years warranty

Specifications

PM-15-5Specifications from the official datasheet (2022-09-20 edition)

OUTPUT

DC VOLTAGE
5V
RATED CURRENT
3A
CURRENT RANGE
0 ~ 3A
RATED POWER
15W
RIPPLE & NOISE (max.)Note.2
80mVp-p
VOLTAGE TOLERANCENote.3
±2.0%
LINE REGULATION
±1.0%
LOAD REGULATION
±1.0%
SETUP, RISE TIME
1000ms, 20ms/230VAC 1000ms, 20ms/115VAC at full load
HOLD UP TIME (Typ.)
100ms/230VAC 24ms/115VAC at full load

INPUT

VOLTAGE RANGE
85 ~ 264VAC 120 ~ 370VDC
FREQUENCY RANGE
47 ~ 440Hz
EFFICIENCY (Typ.)
76%
AC CURRENT (Typ.)
0.35A/115VAC 0.2A/230VAC
INRUSH CURRENT (Typ.)
COLD START 30A/115VAC 50A/230VAC
LEAKAGE CURRENTNote.5
Touch current < 80μ A/264VAC

PROTECTION

OVERLOAD
Above 105% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE
5.75 ~ 6.75V Protection type : Shut off o/p voltage, clamping by zener diode

ENVIRONMENT

WORKING TEMP.
-20 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-40 ~ +85℃, 10 ~ 95% RH
TEMP. COEFFICIENT
±0.03%/℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC (Note 4)

SAFETY STANDARDS
ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1, UL60950-1, EAC TP TC 004 approved
ISOLATION LEVEL
Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE
I/P-O/P:4KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
Compliance to EN55011(CISPR11),EN55032 (CISPR32) Class B, EN61000-3-2,-3, EAC TP TC 020
EMC IMMUNITY
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55035, EN60601-1-2, EN61204-3, medical level, EAC TP TC 020

OTHERS

MTBF
499.7Khrs min. MIL-HDBK-217F (25℃)
DIMENSION
75*53*22.7mm (L*W*H)
PACKING
0.14Kg; 120pcs/17.8Kg/0.94CUFT
Test conditions and notes6
  1. 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
  3. 3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
  5. 5. Touch current was measured from primary input to DC output.
  6. 6. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).

Technical drawings

PM-15-5 Block Diagram
Block Diagram
PM-15-5 Derating Curve
Derating Curve

Downloads

FAQ

Can I use the PM-15-5 for 5V embedded systems or single-board control electronics?

Yes, it provides constant 5V output with up to 3A available. That makes it suitable for 5V embedded or control loads that fit within the 15W output rating.

Is the PM-15-5 a fit for 5V industrial logic or interface devices?

It can power 5V industrial logic or interface electronics where the total draw does not exceed 3A. The unit is rated for 15W total output, so both current and power need to stay within those limits.

Can this model support 5V USB-style or low-voltage automation electronics if the load is fixed at 5V?

It is intended for constant-voltage 5V applications, so it can be used for 5V electronics in automation or embedded equipment when the device accepts a regulated 5V DC supply. The sizing limit is 3A output current and 15W output power.

How much 5V load can the PM-15-5 support?

The output is rated at 5V and 3A, with 15W total power. For compatibility, the connected 5V load should stay at or below 3A and 15W.

Can the PM-15-5 be used in outdoor or unsealed equipment spaces?

The specified operating temperature range is -20°C to +70°C. No IP rating is listed, so the published specs do not show environmental sealing for outdoor, washdown, or wet-area use.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8430994719019, "variants" :[{ "context":"default", "variant_id": 45664579485995, "sku":"PM-15-5", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well

Mean Well PM-15-5 Power Supply 15W 5V 3A

PM-15-5 is the 5V output model of this series: rated current 3A, rated power 15W, typical efficiency 76%.

Features

  • Universal AC input / Full range
  • Protections: Short circuit / Overload / Over voltage
  • Ultra-miniature size, light weight
  • Cooling by free air convection
  • Isolation classⅡ
  • Medical safety approved (2 x MOPP between primary to secondary)
  • No load power consumption<0.5W
  • 100% full load burn-in test
  • Fixed switching frequency at 67KHz
  • High reliability
  • Suitable for BF application with appropriate system consideration
  • 3 years warranty
View product