Overview
MDS03F-12 is the 12V output model of this series: rated power 3W, typical efficiency 80%.
MDS03 and MDD03 series are 3W isolated and regulated module type medical grade DC-DC converter with DIP 24 package. It features international standard pins, a high efficiency up to 81%, wide working temperature range -40~+90℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~36V and 18~75V 4:1 wide input range, and various output voltage, 5V/12V/15V for single output and ±5V/±12V/±15V for dual output. Which are suitable for medical systems, ultra low leakage current.
Features
- DIP 24 package with international standard pinout
- 4:1 wide input range
- Low patient leakage current <5μA
- Wide operating temperature range -40 ~ +90℃
- Medical safety approved (2xMOPP) according to IEC60601-1
- Protections: Short circuit (Continuous) / Overload / UVLO
- 6KVDC or 4.2KVAC high I/O isolation (Reinforced isolation)
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 12V,24V (9 ~ 36V)
- INPUT CURRENT NO LOAD
- 20mA
- INPUT CURRENT FULL LOAD
- 320mA
- VOLTAGE RANGE
- 9 ~ 36Vdc ,
- SURGE VOLTAGE (100ms max.)
- F models : 50Vdc ; G models : 100Vdc
- FILTER
- Pi type
- PROTECTION (Typ.)
- Fuse recommended. F models: 2A slow-blow Type, G models: 1A slow-blow Type
OUTPUT
- OUTPUT VOLTAGE
- 12V
- OUTPUT CURRENT
- 0 ~ 250mA
- EFFICIENCY (TYP.)
- 80%
- CAPACITOR LOAD (MAX.)* For each output
- 47μF
- VOLTAGE ACCURACY
- ±2%
- RATED POWER
- 3W
- RIPPLE & NOISENote.2
- 150mVp-p
- LINE REGULATIONNote.3
- ±0.5%
- LOAD REGULATIONNote.4
- ±0.5%
- SWITCHING FREQUENCY (Typ.)
- 250KHz
PROTECTION
- SHORT CIRCUIT
- Protection type : Continuous, automatic recovery
- OVERLOAD
- 110 ~ 250% rated output power Protection type : Recovers automatically after fault condition is removed
- UNDER VOLTAGE LOCKOUT Start-up voltage
- 8.8Vdc
- UNDER VOLTAGE LOCKOUT Shutdown voltage
- 8Vdc
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +90℃ (Refer to "Derating Curve")
- CASE TEMPERATURE
- +105℃ max.
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.05% / ℃ (0 ~ 71℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 3 ~ 5sec./265℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC ( Note.5)
- SAFETY STANDARDS
- IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:1000M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 2xMOPP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 20pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011 N/A Radiated BS EN/EN55011 Class A without external components - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 ±15KV air Radiated Susceptibility BS EN/EN61000-4-3 10V/m EFT/Bursts BS EN/EN61000-4-4 ±2KV Surge BS EN/EN61000-4-5 ±2KV Line-Line Conducted BS EN/EN61000-4-6 10Vrms(e.m.f.) Magnetic Field BS EN/EN61000-4-8 30A/m
OTHERS
- MTBF
- >454Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 31.8*20.3*12.2mm (1.25*0.8*0.48 inch)
- CASE MATERIAL
- UL94V-0 plastic case
- PACKING
- 17g ; 15pcs/per tube, 750pcs/50 tube max./carton
Test conditions and notes5
- 1.All parameters are specified at normal input(F: 12Vdc, G: 24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load for MDD03.
- 5.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
- Industrial control system
Downloads
FAQ
Can I use the MDS03F-12 to provide 12V for medical control electronics or instrumentation?
Yes, this medical-grade converter delivers a constant 12V output from a 9-36VDC input range. Its 3W power rating and 0.25A output current make it a fit for lower-power 12V electronics rather than heavier 12V loads.
Is the MDS03F-12 suitable for embedded 12V circuits inside test, monitoring, or automation devices?
It is well suited to embedded designs that need a regulated 12V rail and have a wide-input source from 9-36VDC. The 31.8 x 20.3 x 12.2 mm size supports compact layouts in enclosed equipment.
Can this model power low-power 12V subsystems in industrial or medical equipment?
Yes, if the subsystem needs constant 12V and stays within 3W and 0.25A, this model aligns with that requirement. Its medical-grade classification and IEC60601-1 3rd edition (2xMOPP) and EAC approvals may also matter in regulated equipment contexts.
Will the MDS03F-12 handle cold or hot operating conditions, and is it outdoor rated?
The specified operating temperature range is -40C to +90C, with load derating at higher temperatures per the derating curve, so the unit is rated across a broad temperature span. The provided specs do not include an IP rating, so outdoor exposure suitability is not established here.
What are the main differences between MDS03F-12 and its sibling models?
The MDS03F series shown here keeps the same 3W rating and 9-36VDC input range across all variants. The output choices are 5V at 0.6A for MDS03F-05, 12V at 0.25A for this model, and 15V at 0.2A for MDS03F-15.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Please log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.