💡 New range: Boke DALI-2 LED drivers | 10–240 W | 5-year warranty Explore the range

Over 15 years experience of providing support of MEAN WELL Power Supplies

Image shown is a representation only. Exact specifications should be obtained from the product datasheet.

Mean Well MDS03G-05 Medical Grade DC/DC Converter 3W - 18-75V input, 5V output 600mA

5V3WDC/DCMedical Grade
SKU MDS03G-05
Log in to see price Trade pricing & online ordering for account holders
Available to order
|

Overview

MDS03G-05 is the 5V output model of this series: rated power 3W, typical efficiency 80%.

MDS03 and MDD03 series are 3W isolated and regulated module type medical grade DC-DC converter with DIP 24 package. It features international standard pins, a high efficiency up to 81%, wide working temperature range -40~+90℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~36V and 18~75V 4:1 wide input range, and various output voltage, 5V/12V/15V for single output and ±5V/±12V/±15V for dual output. Which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 24 package with international standard pinout
  • 4:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • Protections: Short circuit (Continuous) / Overload / UVLO
  • 6KVDC or 4.2KVAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Specifications

MDS03G-05Specifications from the official datasheet (2025-07-28 edition)

INPUT

INPUT VOLTAGE (RANGE)
Normal 24V,48V (18 ~ 75V)
INPUT CURRENT NO LOAD
15mA
INPUT CURRENT FULL LOAD
160mA
VOLTAGE RANGE
18 ~ 75Vdc
SURGE VOLTAGE (100ms max.)
F models : 50Vdc ; G models : 100Vdc
FILTER
Pi type
PROTECTION (Typ.)
Fuse recommended. F models: 2A slow-blow Type, G models: 1A slow-blow Type

OUTPUT

OUTPUT VOLTAGE
5V
OUTPUT CURRENT
0 ~ 600mA
EFFICIENCY (TYP.)
80%
CAPACITOR LOAD (MAX.)* For each output
330μF
VOLTAGE ACCURACY
±2%
RATED POWER
3W
RIPPLE & NOISENote.2
150mVp-p
LINE REGULATIONNote.3
±0.5%
LOAD REGULATIONNote.4
±0.5%
SWITCHING FREQUENCY (Typ.)
250KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery
OVERLOAD
110 ~ 250% rated output power Protection type : Recovers automatically after fault condition is removed
UNDER VOLTAGE LOCKOUT Start-up voltage
8.8Vdc
UNDER VOLTAGE LOCKOUT Shutdown voltage
8Vdc

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +90℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+105℃ max.
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.05% / ℃ (0 ~ 71℃)
SOLDERING TEMPERATURE
1.5mm from case of 3 ~ 5sec./265℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.5)

SAFETY STANDARDS
IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
WITHSTAND VOLTAGE
I/P-O/P:6KVDC or 4.2KVAC
ISOLATION RESISTANCE
I/P-O/P:1000M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION LEVEL
Primary-secondary: 2xMOPP when system input voltage is with 250VAC, 50/60Hz
ISOLATION CAPACITANCE (Typ.)
20pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011N/A
RadiatedBS EN/EN55011Class A without external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2±15KV air
Radiated SusceptibilityBS EN/EN61000-4-310V/m
EFT/BurstsBS EN/EN61000-4-4±2KV
SurgeBS EN/EN61000-4-5±2KV Line-Line
ConductedBS EN/EN61000-4-610Vrms(e.m.f.)
Magnetic FieldBS EN/EN61000-4-830A/m

OTHERS

MTBF
>454Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
31.8*20.3*12.2mm (1.25*0.8*0.48 inch)
CASE MATERIAL
UL94V-0 plastic case
PACKING
17g ; 15pcs/per tube, 750pcs/50 tube max./carton
Test conditions and notes5
  1. 1.All parameters are specified at normal input(F: 12Vdc, G: 24Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 10% to 100% rated load for MDD03.
  5. 5.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS03G-05 Mechanical Specification
Mechanical Specification
MDS03G-05 Plug Assignment
Plug Assignment
MDS03G-05 MDS03 & MDD03 series
MDS03 & MDD03 series
MDS03G-05 MDS03 & MDD03 series (2)
MDS03 & MDD03 series (2)
MDS03G-05 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
  • Industrial control system

Downloads

FAQ

Can I use the MDS03G-05 to power a 5V embedded controller inside industrial equipment?

Yes, it is a constant-voltage 5V DC/DC converter rated for 3W and up to 0.6A output. That makes it suitable for low-power 5V embedded control circuits that stay within those output limits and can accept the converter's 18-75V input range.

Is the MDS03G-05 a fit for 5V sensor interface or monitoring electronics in automation cabinets?

It can fit low-power 5V sensor interface, monitoring, or logic boards when the load does not exceed 0.6A or 3W. Its compact 31.8 × 20.3 × 12.2 mm size also suits dense control assemblies where space is limited.

Can this converter be used for medical-grade 5V auxiliary power in OEM equipment?

This model is identified as medical grade and is approved to IEC60601-1 3rd edition (2xMOPP) and EAC. For OEM designs that need a regulated 5V auxiliary rail, it is intended for loads up to 0.6A from an 18-75V input source.

How does the MDS03G-05 compare with the other MDS03G models?

All three MDS03G models are 3W medical-grade DC/DC converters with the same 18-75V input range, 31.8 × 20.3 × 12.2 mm dimensions, 0.017 kg weight, -40°C to +90°C operating range, and safety approvals. The main difference is output: the MDS03G-05 provides 5V at 0.6A, the MDS03G-12 provides 12V at 0.25A, and the MDS03G-15 provides 15V at 0.2A.

Is the MDS03G-05 suitable for outdoor or harsh-temperature installations?

Its specified operating temperature is -40°C to +90°C (with derating per the derating curve), so the temperature range is broad for demanding equipment environments. However, no IP rating is listed, so these specs do not establish outdoor, washdown, or exposed-weather suitability on their own.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8442610352427, "variants" :[{ "context":"default", "variant_id": 45711611625771, "sku":"MDS03G-05", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS03G-05 Medical Grade DC/DC Converter 3W - 18-75V input, 5V output 600mA
Mean Well

Mean Well MDS03G-05 Medical Grade DC/DC Converter 3W - 18-75V input, 5V output 600mA

MDS03G-05 is the 5V output model of this series: rated power 3W, typical efficiency 80%.

MDS03 and MDD03 series are 3W isolated and regulated module type medical grade DC-DC converter with DIP 24 package. It features international standard pins, a high efficiency up to 81%, wide working temperature range -40~+90℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~36V and 18~75V 4:1 wide input range, and various output voltage, 5V/12V/15V for single output and ±5V/±12V/±15V for dual output. Which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 24 package with international standard pinout
  • 4:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • Protections: Short circuit (Continuous) / Overload / UVLO
  • 6KVDC or 4.2KVAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
  • Industrial control system
View product