Mean Well RPT-75D Triple Output Medical Grade 6A

5" x 3" compact sizemedical safety approvedextremely low leakage currenttriple outputlifetime > 140K hours
SKU RPT-75D
Log in to see price Trade pricing & online ordering for account holders

Request a quote for one or many products — we'll reply with pricing and availability.

Available to order
|

Overview

RPT-75D is the 5V 24V 12V output model of this series: rated current 5A 1.5A 1A, rated power 73W, typical efficiency 79%.

RPT-75 is a 72W highly reliable PCB type medical power supply with a high power density on the 5” by 3” footprint. It accepts 90~264VAC input and offers triple output voltages . RPT-75 is able to be used for Class Ⅰ system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011.a

Features

  • 5"×3" compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • 72W convection, 100W force air
  • EMI class B for class Ⅰ configuration
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 140K hours
  • 3 years warranty

Specifications

RPT-75DSpecifications from the official datasheet (2024-11-22 edition)

OUTPUT

OUTPUT NUMBER
CH1 CH2 CH3
DC VOLTAGE
5V 24V 12V
RATED CURRENT
5A 1.5A 1A
CURRENT RANGE
0.6 ~ 7A 0.1 ~ 2A 0.1 ~ 1A
RATED POWER
73W
PEAK LOAD (23.5CFM)
95W
RIPPLE & NOISE (max.)Note.2
80mVp-p 200mVp-p 120mVp-p
VOLTAGE ADJ. RANGE
CH1:4.75 ~ 5.5V
VOLTAGE TOLERANCENote.3
±2.0% ±8.0% ±8.0%
LINE REGULATION
±0.5% ±1.0% ±1.0%
LOAD REGULATION
±1.5% ±3.0% ±3.0%
SETUP, RISE TIME
500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
HOLD UP TIME (Typ.)
90ms/230VAC 20ms/115VAC at full load

INPUT

VOLTAGE RANGE
90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY(Typ.)
79%
AC CURRENT (Typ.)
1.5A/115VAC 1A/230VAC
INRUSH CURRENT (Typ.)
COLD START 25A/115VAC 50A/230VAC
LEAKAGE CURRENTNote.4
Earth leakage current < 150μA/264VAC , Touch current < 100μA/264VAC

PROTECTION

OVERLOAD
140 ~ 180% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE
Ch1: 5.7 ~ 6.8V Protection type : Shut down o/p voltage, re-power on to recover

ENVIRONMENT

WORKING TEMP.
-20 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
±0.03%/℃ (0 ~ 45℃)
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDENote.5
3000 meters

SAFETY & EMC (Note 8)

SAFETY STANDARDS
IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2 CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL
Primary-Secondary:2xMOPP, Primary-Earth:1xMOPP
WITHSTAND VOLTAGE
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION RESISTANCE
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
ParameterStandardTest Level
Conducted emissionBS EN/EN55011 (CISPR11)Class B
Radiated emissionBS EN/EN55011 (CISPR11)Class B
Harmonic currentBS EN/EN61000-3-2Class A
Voltage flickerBS EN/EN61000-3-3
EMC IMMUNITY

BS EN/EN55035,BS EN/EN60601-1-2

ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 4, 15KV air; Level 4, 8KV contact
RF field susceptibilityBS EN/EN61000-4-3Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT burstsBS EN/EN61000-4-4Level 3, 2KV
Surge susceptibilityBS EN/EN61000-4-5Level 4, 4KV/Line-FG; 2KV/Line-Line
Conducted susceptibilityBS EN/EN61000-4-6Level 3, 10V
Magnetic field immunityBS EN/EN61000-4-8Level 4, 30A/m
Voltage dip, interruptionBS EN/EN61000-4-11100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods

OTHERS

MTBF
2305.8K hrs min. Telcordia SR-332 (Bellcore) ; 521.3K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (L*W*H)
127*76.2*31mm or 5" * 3" *1.22" inch
PACKING
0.25Kg; 63pcs/17.3Kg/1.28CUFT
Test conditions and notes8
  1. 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
  3. 3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. 4. Touch current was measured from primary input to DC output.
  5. 5. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  6. 6. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.
  7. 7. Heat Sink HS1,HS2,HS3 can not be shorted.
  8. 8. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

Technical drawings

RPT-75D Block Diagram
Block Diagram
RPT-75D Derating Curve
Derating Curve
RPT-75D Mechanical Specification
Mechanical Specification

Typical applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

Downloads

FAQ

Can I use the RPT-75D for a 5V medical or laboratory device?

This model has a 5V main output rated for 5A and a total power rating of 73W. Its features list medical safety approval and extremely low leakage current, which makes it relevant for medical or lab electronics that need a compact triple output supply. The dimensions are 127.0 x 76.2 x 31.0 mm.

Is the RPT-75D suitable for embedded control electronics with a 5V primary rail?

It can fit embedded control hardware that runs from a regulated 5V rail and needs up to 5A on that main output. The supply operates in constant voltage mode and is rated for 73W total. Its triple output format may also help in designs that need more than one rail from a single unit.

Would the RPT-75D work in a compact industrial control enclosure?

The power level is 73W, and the form factor is 127.0 x 76.2 x 31.0 mm with a weight of 0.25 kg. That makes it more suitable for compact electronics enclosures than for bulkier high-power systems. The universal input range is 90-264VAC.

Is the RPT-75D rated for outdoor use or only temperature-tested?

The specified operating temperature range is -20C to +70C. No IP rating is listed in the provided specifications, so these details do not establish outdoor or wet-environment suitability. Based on the published data, the environmental guidance here is limited to temperature range.

How does the RPT-75D compare with the other RPT siblings listed here?

The listed family spans from 3W to 75W. Specifically, RPT-160A, RPT-160B, RPT-160C, and RPT-160D are 3W; RPT-6003 is 46.5W; RPT-60A is 46.5W with 5V and 4A; RPT-60B is 50W with 5V and 4A; RPT-60C is 50W with 5V and 4A; RPT-60D is 46.5W; RPT-65E is 62.5W; RPT-65E-C is 62.5W; RPT-65F is 62.5W; RPT-65G is 62.5W; RPT-65G-C is 62.5W; RPT-7503 is 61.8W; RPT-75A is 68.5W with 5V and 6A; RPT-75B is 72W with 5V and 6A; and RPT-75C is 72W with 5V and 6A. RPT-65F-C is included in the sibling list without a wattage in the supplied data.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8431176646955, "variants" :[{ "context":"default", "variant_id": 45665682063659, "sku":"RPT-75D", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well RPT-75D Triple Output Medical Grade 6A
Mean Well

Mean Well RPT-75D Triple Output Medical Grade 6A

RPT-75D is the 5V 24V 12V output model of this series: rated current 5A 1.5A 1A, rated power 73W, typical efficiency 79%.

RPT-75 is a 72W highly reliable PCB type medical power supply with a high power density on the 5” by 3” footprint. It accepts 90~264VAC input and offers triple output voltages . RPT-75 is able to be used for Class Ⅰ system design. The extremely low leakage current is less than 150μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011.a

Features

  • 5"×3" compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • 72W convection, 100W force air
  • EMI class B for class Ⅰ configuration
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • Lifetime > 140K hours
  • 3 years warranty

Typical applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices
View product