Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
MDD03G-05 is the ±5V output model of this series: rated power 3W, typical efficiency 80%.
MDS03 and MDD03 series are 3W isolated and regulated module type medical grade DC-DC converter with DIP 24 package. It features international standard pins, a high efficiency up to 81%, wide working temperature range -40~+90℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~36V and 18~75V 4:1 wide input range, and various output voltage, 5V/12V/15V for single output and ±5V/±12V/±15V for dual output. Which are suitable for medical systems, ultra low leakage current.
Features
- DIP 24 package with international standard pinout
- 4:1 wide input range
- Low patient leakage current <5μA
- Wide operating temperature range -40 ~ +90℃
- Medical safety approved (2xMOPP) according to IEC60601-1
- Protections: Short circuit (Continuous) / Overload / UVLO
- 6KVDC or 4.2KVAC high I/O isolation (Reinforced isolation)
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 24V,48V (18 ~ 75V)
- INPUT CURRENT NO LOAD
- 15mA
- INPUT CURRENT FULL LOAD
- 158mA
- VOLTAGE RANGE
- 18 ~ 75Vdc
- SURGE VOLTAGE (100ms max.)
- F models : 50Vdc ; G models : 100Vdc
- FILTER
- Pi type
- PROTECTION (Typ.)
- Fuse recommended. F models: 2A slow-blow Type, G models: 1A slow-blow Type
OUTPUT
- OUTPUT VOLTAGE
- ±5V
- OUTPUT CURRENT
- ±30 ~ 300mA
- EFFICIENCY (TYP.)
- 80%
- CAPACITOR LOAD (MAX.)* For each output
- *220μF
- VOLTAGE ACCURACY
- ±2%
- RATED POWER
- 3W
- RIPPLE & NOISENote.2
- 150mVp-p
- LINE REGULATIONNote.3
- ±0.5%
- LOAD REGULATIONNote.4
- ±0.5%
- SWITCHING FREQUENCY (Typ.)
- 250KHz
PROTECTION
- SHORT CIRCUIT
- Protection type : Continuous, automatic recovery
- OVERLOAD
- 110 ~ 250% rated output power Protection type : Recovers automatically after fault condition is removed
- UNDER VOLTAGE LOCKOUT Start-up voltage
- 8.8Vdc
- UNDER VOLTAGE LOCKOUT Shutdown voltage
- 8Vdc
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +90℃ (Refer to "Derating Curve")
- CASE TEMPERATURE
- +105℃ max.
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.05% / ℃ (0 ~ 71℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 3 ~ 5sec./265℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC ( Note.5)
- SAFETY STANDARDS
- IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:1000M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 2xMOPP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 20pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011 N/A Radiated BS EN/EN55011 Class A without external components - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 ±15KV air Radiated Susceptibility BS EN/EN61000-4-3 10V/m EFT/Bursts BS EN/EN61000-4-4 ±2KV Surge BS EN/EN61000-4-5 ±2KV Line-Line Conducted BS EN/EN61000-4-6 10Vrms(e.m.f.) Magnetic Field BS EN/EN61000-4-8 30A/m
OTHERS
- MTBF
- >454Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 31.8*20.3*12.2mm (1.25*0.8*0.48 inch)
- CASE MATERIAL
- UL94V-0 plastic case
- PACKING
- 17g ; 15pcs/per tube, 750pcs/50 tube max./carton
Test conditions and notes5
- 1.All parameters are specified at normal input(F: 12Vdc, G: 24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load for MDD03.
- 5.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
- Industrial control system
Downloads
FAQ
Can I use the MDD03G-05 in a medical monitor or patient-adjacent embedded system that needs 5V rails?
This model is a 3W constant-voltage DC/DC converter with ±5V output at up to 300mA per output, so it fits low-power 5V medical or embedded subcircuits that stay within that output limit. Its medical-grade positioning and IEC60601-1 3rd edition (2xMOPP) and EAC approvals make it appropriate for regulated equipment designs where those approvals matter.
Is the MDD03G-05 suitable for industrial control electronics that need isolated low-power 5V conversion?
It can fit compact industrial control or automation electronics that need ±5V rails and draw no more than 300mA per output from a 3W converter. The 18-75V input range also suits systems built around a wide nominal DC bus within that stated range.
Can this converter power 5V logic in embedded automation or data acquisition boards?
Yes, it is aligned with low-power 5V logic sections in embedded or data acquisition designs, provided the load stays within ±5V and 300mA per output. The small 31.8 × 20.3 × 12.2 mm package also helps where board or enclosure space is limited.
Which MDD03G model should I choose: 5V, 12V, or 15V?
Within this 3W series, the MDD03G-05 provides ±5V at 300mA, the MDD03G-12 provides ±12V at 125mA, and the MDD03G-15 provides ±15V at 100mA per output. All three share the same 18-75V input range, 31.8 × 20.3 × 12.2 mm dimensions, 0.017 kg weight, and -40°C to +90°C operating temperature range.
Is the MDD03G-05 suitable for outdoor or harsh-temperature installations?
The operating temperature range is -40°C to +90°C (with derating per the derating curve), so the temperature spec supports use across a wide range of enclosed equipment environments. No IP rating is listed in the provided specs, so outdoor exposure or washdown suitability cannot be confirmed from this data alone.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other MDD03 Models
5 other models| Model | Power | Voltage | Current | IP Rating | |
|---|---|---|---|---|---|
| Mean Well MDD03G-15 Medical Grade DC/DC Converter 3W - 18-75V input, 15 and -15V output 200mA | 3W | 15V | 0.2A | IP20 | View |
| Mean Well MDD03G-12 Medical Grade DC/DC Converter 3W - 18-75V input, 12 and -12V output 250mA | 3W | 12V | 0.25A | IP20 | View |
| Mean Well MDD03F-15 Medical Grade DC/DC Converter 3W - 9-36V input, 15 and -15V output 200mA | 3W | 15V | 0.2A | IP20 | View |
| Mean Well MDD03F-12 Medical Grade DC/DC Converter 3W - 9-36V input, 12 and -12V output 250mA | 3W | 12V | 0.25A | IP20 | View |
| Mean Well MDD03F-05 Medical Grade DC/DC Converter 3W - 9-36V input, 5 and -5V output 600mA | 3W | 5V | 0.6A | IP20 | View |
Please log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.