💡 New range: Boke DALI-2 LED drivers | 10–240 W | 5-year warranty Explore the range

Over 15 years experience of providing support of MEAN WELL Power Supplies

Image shown is a representation only. Exact specifications should be obtained from the product datasheet.

Mean Well MDS20A-05 DC-DC Converter - 20W - 9~18V in 5V out 4A

SKU MDS20A-05
Log in to see price Trade pricing & online ordering for account holders
Available to order
|

Overview

MDS20A-05 is the 5V output model of this series: rated power 20W, typical efficiency 86%.

MDS20 series is 20W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 90%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Specifications

MDS20A-05Specifications from the official datasheet (2024-04-30 edition)

INPUT

INPUT VOLTAGE (RANGE)
Normal 12V (9 ~ 18V)
INPUT CURRENT NO LOAD
10mA
INPUT CURRENT FULL LOAD
1950mA
VOLTAGE RANGE
9~18Vdc
SURGE VOLTAGE (100ms max.)
12Vin models : 25Vdc, 24Vin models : 50Vdc, 48Vin models : 100Vdc
FILTER
Pi type
PROTECTION
Fuse recommended. 12Vin models: 5A delay time Type, 24Vin models: 3A delay time Type, 48Vin models: 2A delay time Type

OUTPUT

OUTPUT VOLTAGE
5V
OUTPUT CURRENT
4000mA
EFFICIENCY (Typ.)
86%
CAPACITOR LOAD (MAX.)
6800μF
VOLTAGE ACCURACY
±1%
RATED POWER
20W
RIPPLE & NOISENote.2
60mVp-p
LINE REGULATIONNote.3
±0.5%
LOAD REGULATIONNote.4
±0.5%
SWITCHING FREQUENCY (Typ.)
12Vin/24Vin:250KHz, 48Vin:300KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery
OVERLOAD
110 ~ 180% rated output power Protection type : Recovers automatically after fault condition is removed
UNDER VOLTAGE LOCKOUT
12Vin: 7.5Vdc, 24Vin: 15Vdc, 48Vin: 33Vdc

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +90℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+105℃ max.
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.03% / ℃ (0 ~ 70℃)
SOLDERING TEMPERATURE
1.5mm from case of 1 ~ 3sec./260℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.6)

SAFETY STANDARDS
IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
WITHSTAND VOLTAGE
I/P-O/P 4KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION CAPACITANCE (Typ.)
100pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011Class A without external components
RadiatedBS EN/EN55011Class A without external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 1, ±15KV air, ±8KV contact
Radiated SusceptibilityBS EN/EN61000-4-3Level 1, 10V/m
EFT/BurstsBS EN/EN61000-4-4Level 1, ±2KV
SurgeBS EN/EN61000-4-5Level 1, ±1KV Line-Line, ±2KV Line-GND
ConductedBS EN/EN61000-4-6Level 1, 10Vrms(e.m.f.)

OTHERS

MTBF
1016Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
50.8*25.4*12mm (2*1*0.47 inch)
CASE MATERIAL
UL94V-0 plastic case
PACKING
30g; 18pcs/per tube, 432pcs/24 tube max./carton
Test conditions and notes6
  1. 1.All parameters are specified at normal input(A:12Vdc, B:24Vdc, C:48Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 0% to 100% rated load.
  5. 5.2xMOPP base on a working voltage of 250Vac between Primary to Secondary.
  6. 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS20A-05 Pin Configuration
Pin Configuration
MDS20A-05 Derating Curve
Derating Curve
MDS20A-05 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment

Downloads

FAQ

Can I use the MDS20A-05 to derive 5V logic power from a 12V or 24V-class mobile battery system?

It is designed for a 9-18V input range, so it fits 12V-class DC systems but not inputs above 18V. The output is 5V constant voltage with a 20W rating and 4A output current. That makes it appropriate for heavier 5V embedded, controller, or communication loads within those limits.

Is the MDS20A-05 a good choice for powering 5V embedded computers or industrial interface modules?

Yes, this model provides a regulated 5V output and offers up to 20W, which is useful for 5V embedded electronics and interface hardware that draw more current than small signal converters can provide. The rated output current is 4A. It is intended for systems with a 9-18V input source.

Can this converter support 5V loads in vehicle, telecom, or automation cabinets with limited space?

Its 50.8 × 25.4 × 12.0 mm size and 0.03 kg weight suit compact electronics assemblies and tight enclosure layouts. Electrically, it converts a 9-18V input to a 5V constant-voltage output at up to 20W. That profile matches space-constrained low-voltage subsystems better than higher-power chassis supplies.

What is the difference between the MDS20A-05 and the other MDS20A versions?

The MDS20A family keeps the same 20W power class but changes the output voltage and rated current. The MDS20A-05 is 5V at 4A, the MDS20A-12 is 12V at 1.67A, the MDS20A-15 is 15V at 1.33A, and the MDS20A-24 is 24V at 833mA. This model is the 5V option in that series.

Can the MDS20A-05 be used in outdoor equipment or high-temperature enclosures?

The operating temperature range is -40°C to +90°C, with output derating at higher ambient temperatures per the datasheet derating curve. No IP rating is listed in the provided specs, so it is better suited to protected indoor panels, cabinets, or sealed enclosures than direct outdoor exposure. The temperature rating does not by itself indicate weather resistance.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8442613170475, "variants" :[{ "context":"default", "variant_id": 45711622144299, "sku":"MDS20A-05", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS20A-05 DC-DC Converter - 20W - 9~18V in 5V out 4A
Mean Well

Mean Well MDS20A-05 DC-DC Converter - 20W - 9~18V in 5V out 4A

MDS20A-05 is the 5V output model of this series: rated power 20W, typical efficiency 86%.

MDS20 series is 20W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 90%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
View product