Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
MDS20B-05 is the 5V output model of this series: rated power 20W, typical efficiency 86%.
MDS20 series is 20W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 90%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.
Features
- DIP 2"x1" package with international standard pinout
- 2:1 wide input range
- Low patient leakage current <5μA
- Wide operating temperature range -40 ~ +90℃
- Medical safety approved (2xMOPP) according to IEC60601-1
- No minimum load required
- Protections: Short circuit (Continuous) / Overload / Input under voltage
- 4000VAC high I/O isolation (Reinforced isolation)
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 24V (18 ~ 36V)
- INPUT CURRENT NO LOAD
- 7mA
- INPUT CURRENT FULL LOAD
- 1000mA
- VOLTAGE RANGE
- 18~36Vdc
- SURGE VOLTAGE (100ms max.)
- 12Vin models : 25Vdc, 24Vin models : 50Vdc, 48Vin models : 100Vdc
- FILTER
- Pi type
- PROTECTION
- Fuse recommended. 12Vin models: 5A delay time Type, 24Vin models: 3A delay time Type, 48Vin models: 2A delay time Type
OUTPUT
- OUTPUT VOLTAGE
- 5V
- OUTPUT CURRENT
- 4000mA
- EFFICIENCY (Typ.)
- 86%
- CAPACITOR LOAD (MAX.)
- 6800μF
- VOLTAGE ACCURACY
- ±1%
- RATED POWER
- 20W
- RIPPLE & NOISENote.2
- 60mVp-p
- LINE REGULATIONNote.3
- ±0.5%
- LOAD REGULATIONNote.4
- ±0.5%
- SWITCHING FREQUENCY (Typ.)
- 12Vin/24Vin:250KHz, 48Vin:300KHz
PROTECTION
- SHORT CIRCUIT
- Protection type : Continuous, automatic recovery
- OVERLOAD
- 110 ~ 180% rated output power Protection type : Recovers automatically after fault condition is removed
- UNDER VOLTAGE LOCKOUT
- 12Vin: 7.5Vdc, 24Vin: 15Vdc, 48Vin: 33Vdc
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +90℃ (Refer to "Derating Curve")
- CASE TEMPERATURE
- +105℃ max.
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.03% / ℃ (0 ~ 70℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC ( Note.6)
- SAFETY STANDARDS
- IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
- WITHSTAND VOLTAGE
- I/P-O/P 4KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION CAPACITANCE (Typ.)
- 100pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011 Class A without external components Radiated BS EN/EN55011 Class A without external components - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 1, ±15KV air, ±8KV contact Radiated Susceptibility BS EN/EN61000-4-3 Level 1, 10V/m EFT/Bursts BS EN/EN61000-4-4 Level 1, ±2KV Surge BS EN/EN61000-4-5 Level 1, ±1KV Line-Line, ±2KV Line-GND Conducted BS EN/EN61000-4-6 Level 1, 10Vrms(e.m.f.)
OTHERS
- MTBF
- 1016Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 50.8*25.4*12mm (2*1*0.47 inch)
- CASE MATERIAL
- UL94V-0 plastic case
- PACKING
- 30g; 18pcs/per tube, 432pcs/24 tube max./carton
Test conditions and notes6
- 1.All parameters are specified at normal input(A:12Vdc, B:24Vdc, C:48Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 0% to 100% rated load.
- 5.2xMOPP base on a working voltage of 250Vac between Primary to Secondary.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
Downloads
FAQ
Can I use the MDS20B-05 to power 5V PLC or industrial control electronics from an 18-36V bus?
This model is intended for constant voltage output at 5V from an 18-36V input range. With 20W total power and up to 4A output current, it fits 5V industrial control, embedded, and automation electronics that stay within those limits.
Is the MDS20B-05 suitable for 5V embedded systems inside automation equipment?
It can be a fit for embedded boards, logic modules, and 5V control circuits that need a regulated 5V output. The unit provides up to 20W and 4A, so the connected 5V load needs to remain within that capacity.
Can this converter be used for 5V monitoring or sensor interfaces in industrial cabinets?
Yes, it is designed as a 5V constant voltage converter with an 18-36V input range, which suits many cabinet-based monitoring and interface loads. Its compact 50.8 × 25.4 × 12.0 mm size also helps where control enclosure space is limited.
How does the MDS20B-05 compare with the other MDS20B output options?
All MDS20B models in this group are rated for 20W and share the same 18-36V input range. The difference is output: the MDS20B-05 is 5V at 4A, the MDS20B-12 is 12V at 1.67A, the MDS20B-15 is 15V at 1.33A, and the MDS20B-24 is 24V at 833mA.
Is the MDS20B-05 suitable for outdoor or harsh-temperature locations?
The operating temperature range is -40°C to +90°C, with output derating at higher ambient temperatures per the datasheet derating curve. No IP rating is listed in the provided specs, so these specs do not indicate suitability for exposed outdoor or wet locations.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.