Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
MDD01L-05 is the ±5V output model of this series: rated power 1W, typical efficiency 79%.
MDS01 and MDD01 series are 1W isolated and unregulated module type medical grade DC-DC converter with SIP7 package. It features international standard pins, a high efficiency up to 83%, wide working temperature range -40~+85℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 3.3V/ 5V/ 12 V/15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems , ultra low leakage current.
Features
- SIP7 package with international standard pinout
- Operating temperature range -40 ~ +85℃
- Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
- Low patient leakage current <2μA
- Protection: Short circuit(3 second max.)
- 6KVDC or 4.2KVAC hight I/O isolation (Reinforced isolation)
- Low cost
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 5V (4.5 ~ 5.5V)
- VOLTAGE RANGE
- 4.5 ~ 5.5Vdc
- INPUT CURRENT (NO LOAD)
- 25mA
- INPUT CURRENT (FULL LOAD)
- 260mA
- SURGE VOLTAGE (100ms max.)
- 5Vin models : 9Vdc 12Vin models : 18Vdc 24Vin models : 30Vdc
- FILTER
- Internal capacitor
- PROTECTION
- Fuse recommended. 500mA Slow-Blow Type for all models
OUTPUT
- OUTPUT VOLTAGE
- ±5V
- OUTPUT CURRENT
- ±10 ~ 100mA
- EFFICIENCY (TYP.)
- 79%
- CAPACITOR LOAD (MAX.)* For each output
- *470μF
- VOLTAGE ACCURACY
- ±5.0%
- RATED POWER
- 1W
- RIPPLE & NOISENote.2
- 75mVp-p
- LINE REGULATIONNote.3
- 1.2% for 1% input variation
- LOAD REGULATIONNote.4
- ±10%
- SWITCHING FREQUENCY (Typ.)
- 100KHz
PROTECTION
- SHORT CIRCUIT
- 3 second max.
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +85℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.02% / ℃ (0 ~ 85℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC
- SAFETY STANDARDS
- UL60601-1, EAC TP TC 020/2011 approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 5pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011(CISPR11) Class B Radiated BS EN/EN55011(CISPR11) Class B - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 2, ±8KV contact
OTHERS
- MTBF
- 3500Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 19.5*9.8*12.5mm (0.77*0.386*0.492 inch)
- CASE MATERIAL
- Non-Conductive black plastic (UL 94V-0 rated)
- PACKING
- 4.2g ; 25pcs/per tube, 3000pcs/120 tube/per carton
Test conditions and notes6
- 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load.
- 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
Downloads
Certificates4
Test reports1
FAQ
Can I use the MDD01L-05 for a low-power 5V automation control board?
Yes, if the board needs a ±5V dual rail. The unit provides a fixed, unregulated ±5V output at 1W from a 4.5~5.5VDC input, which makes it a better fit for small electronics than for larger automation loads.
Is this converter appropriate for 5V industrial communication or monitoring circuits?
It can be used where a low-power ±5V supply is needed for monitoring, signaling, or communication subcircuits in industrial equipment. The key constraint is its 1W power rating (up to 100mA per output), so it should be reserved for compact electronic sections instead of full system loads.
Can the MDD01L-05 support 5V embedded peripherals in an OEM device?
It is a reasonable choice for low-power ±5V embedded peripherals, controller support circuits, or interface boards in OEM equipment. The converter provides a fixed ±5V output and operates across -40°C to +85°C (with load derating at higher temperatures, see the derating curve), which is useful for electronics deployed in varied operating environments.
Is the MDD01L-05 suitable for outdoor installations?
The operating temperature range is -40°C to +85°C (with load derating at higher temperatures, see the derating curve), so the temperature specification is broad. However, no IP rating is listed in the provided data, so these specs alone do not confirm outdoor exposure or weather-resistant use.
What are the main output choices in the MDD01L family compared with the MDD01L-05?
The 1W MDD01L siblings are MDD01L-09 at ±9V up to 56mA, MDD01L-12 at ±12V up to 42mA, and MDD01L-15 at ±15V up to 34mA. This model is the ±5V version rated at 1W, with up to 100mA per output.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other MDD01 Models
17 other modelsPlease log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.