💡 New range: Boke DALI-2 LED drivers | 10–240 W | 5-year warranty Explore the range

Over 15 years experience of providing support of MEAN WELL Power Supplies

Image shown is a representation only. Exact specifications should be obtained from the product datasheet.

Mean Well MDS06F-12 Medical Grade DC/DC Converter 6W - 9-36V input, 12V output 500mA

12V6WDC/DCMedical Grade
SKU MDS06F-12
Log in to see price Trade pricing & online ordering for account holders
Available to order
|

Overview

MDS06F-12 is the 12V output model of this series: rated power 6W, typical efficiency 81%.

MDS06 and MDD06 series are 6W isolated and regulated module type medical grade DC-DC converter with DIP 24 package. It features international standard pins, a high efficiency up to 84%, wide working temperature range -40~+90℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~36V and 18~75V 4:1 wide input range, and various output voltage, 5V/12V/15V for single output and ±5V/±12V/±15V for dual output. Which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 24 package with international standard pinout
  • 4:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • Protections: Short circuit (Continuous) / Overload / UVLO
  • 6KVDC or 4.2KVAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Specifications

MDS06F-12Specifications from the official datasheet (2025-07-28 edition)

INPUT

INPUT VOLTAGE (RANGE)
Normal 12V,24V (9 ~ 36V)
INPUT CURRENT NO LOAD
22mA
INPUT CURRENT FULL LOAD
620mA
VOLTAGE RANGE
9 ~ 36Vdc ,
SURGE VOLTAGE (100ms max.)
F models : 50Vdc ; G models : 100Vdc
FILTER
Pi type
PROTECTION (Typ.)
Fuse recommended. F models: 2A slow-blow Type, G models: 1A slow-blow Type

OUTPUT

OUTPUT VOLTAGE
12V
OUTPUT CURRENT
0 ~ 500mA
EFFICIENCY (TYP.)
81%
CAPACITOR LOAD (MAX.)* For each output
120μF
VOLTAGE ACCURACY
±2%
RATED POWER
6W
RIPPLE & NOISENote.2
100mVp-p
LINE REGULATIONNote.3
±0.5%
LOAD REGULATIONNote.4
±0.5%
SWITCHING FREQUENCY (Typ.)
280KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery
OVERLOAD
110 ~ 250% rated output power Protection type : Recovers automatically after fault condition is removed
UNDER VOLTAGE LOCKOUT Start-up voltage
8.8Vdc
UNDER VOLTAGE LOCKOUT Shutdown voltage
8Vdc

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +90℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+105℃ max.
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.05% / ℃ (0 ~ 71℃)
SOLDERING TEMPERATURE
1.5mm from case of 3 ~ 5sec./265℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.5)

SAFETY STANDARDS
IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
WITHSTAND VOLTAGE
I/P-O/P:6KVDC or 4.2KVAC
ISOLATION RESISTANCE
I/P-O/P:1000M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION CAPACITANCE (Typ.)
20pF
ISOLATION LEVEL
Primary-secondary: 2xMOPP when system input voltage is with 250VAC, 50/60Hz
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011N/A
RadiatedBS EN/EN55011Class A without external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2±15KV air
Radiated SusceptibilityBS EN/EN61000-4-310V/m
EFT/BurstsBS EN/EN61000-4-4±2KV
SurgeBS EN/EN61000-4-5±2KV Line-Line
ConductedBS EN/EN61000-4-610Vrms(e.m.f.)
Magnetic FieldBS EN/EN61000-4-830A/m

OTHERS

MTBF
>454Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
31.8*20.3*12.2mm (1.25*0.8*0.48 inch)
CASE MATERIAL
UL94V-0 plastic case
PACKING
17g ; 15pcs/per tube, 750pcs/50 tube max./carton
Test conditions and notes5
  1. 1.All parameters are specified at normal input(F: 12Vdc, G: 24Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 10% to 100% rated load for MDD06.
  5. 5.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS06F-12 Mechanical Specification
Mechanical Specification
MDS06F-12 Plug Assignment
Plug Assignment
MDS06F-12 MDS06 & MDD06 series
MDS06 & MDD06 series
MDS06F-12 MDS06 & MDD06 series (2)
MDS06 & MDD06 series (2)
MDS06F-12 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
  • Industrial control system

Downloads

FAQ

Can I use the MDS06F-12 for 12V industrial control or instrumentation electronics?

Yes, for 12V loads that stay within 0.5A and 6W. That rating suits low-power control, instrumentation, or interface electronics fed from a 9-36V input source.

Is the MDS06F-12 suitable for small 12V automation modules inside mobile or panel systems?

It can fit small 12V automation modules when the total load remains at or below 0.5A. The 9-36V input range is also useful where the available DC bus is lower than the 18-75V range of the G models in this series.

Can this model provide a 12V auxiliary supply for medical or embedded OEM designs?

Yes, for low-power 12V auxiliary rails up to 6W. It is a medical-grade converter approved to IEC60601-1 3rd edition and EAC, with 2xMOPP isolation and a constant-voltage 12V output.

What are the output differences across the MDS06F family?

The MDS06F family shares the same 6W power rating, 9-36V input range, 31.8 × 20.3 × 12.2 mm size, 0.017 kg weight, -40°C to +90°C operating range, and IEC60601-1 / EAC approvals. Output options are 5V at 1.2A for MDS06F-05, 12V at 0.5A for MDS06F-12, and 15V at 0.4A for MDS06F-15.

Is the MDS06F-12 appropriate for exposed outdoor conditions?

The published operating temperature range is -40°C to +90°C, with output derating at higher ambient temperatures per the derating curve. However, no IP rating is listed, so the specs do not confirm outdoor, rain, or dust-exposed use by themselves.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8442610876715, "variants" :[{ "context":"default", "variant_id": 45711613493547, "sku":"MDS06F-12", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS06F-12 Medical Grade DC/DC Converter 6W - 9-36V input, 12V output 500mA
Mean Well

Mean Well MDS06F-12 Medical Grade DC/DC Converter 6W - 9-36V input, 12V output 500mA

MDS06F-12 is the 12V output model of this series: rated power 6W, typical efficiency 81%.

MDS06 and MDD06 series are 6W isolated and regulated module type medical grade DC-DC converter with DIP 24 package. It features international standard pins, a high efficiency up to 84%, wide working temperature range -40~+90℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~36V and 18~75V 4:1 wide input range, and various output voltage, 5V/12V/15V for single output and ±5V/±12V/±15V for dual output. Which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 24 package with international standard pinout
  • 4:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • Protections: Short circuit (Continuous) / Overload / UVLO
  • 6KVDC or 4.2KVAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
  • Industrial control system
View product