Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
MDS01M-12N is the 12V output model of this series: rated power 1W, typical efficiency 81%.
MDS01-N and MDD01-N series are 1W isolated and unregulated module type medical grade DC-DC converter with SIP7 package. It features international standard pins, a high efficiency up to 84%, wide working temperature range -40~+95℃, 6KVDC I/P-O/P hi-ght isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 3.3V/5V/12V/15V for single output and ±5V/±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current or industrial control, telecommunication field, distributed power architecture, and so on.
Features
- SIP7 package with international standard pinout
- Operating temperature range -40 ~ +95℃
- Comply with IEC/UL62368-1 and LVD IEC60601-1
- Medical safety approved (1xMOPP/2xMOOP) according to IEC60601-1
- Low patient leakage current <2μA
- Protections: Short circuit (Continuous)
- 6KVDC hi-ght I/O isolation (Reinforced insulation)
- Low cost
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 12V (10.8 ~ 13.2V)
- VOLTAGE RANGE
- 10.8 ~ 13.2Vdc
- INPUT CURRENT (NO LOAD)
- 15mA
- INPUT CURRENT (FULL LOAD)
- 103mA
- FILTER
- Internal capacitor
- PROTECTION
- Fuse recommended. 500mA Slow-Blow Type for all models
OUTPUT
- OUTPUT VOLTAGE
- 12V
- OUTPUT CURRENT
- 9 ~ 84mA
- EFFICIENCY(Typ.)
- 81%
- CAPACITOR LOAD(Max.)
- 220μF
- VOLTAGE ACCURACY
- ±5.0%
- RATED POWER
- 1W
- RIPPLE & NOISENote.2
- 100mVp-p
- LINE REGULATIONNote.3
- 1.2% for 1% input variation
- LOAD REGULATIONNote.4
- ±15% for 3.3Vo, 5Vo ; ±10% for others
- SWITCHING FREQUENCY(Typ.)
- 30KHz
PROTECTION
- SHORT CIRCUIT
- Protection type : Continuous, automatic recovery
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +95℃ (Refer to "Derating Curve")
- CASE TEMPERATURE
- +105℃
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.02% / ℃ (0 ~ 90℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC
- SAFETY STANDARDS
- IEC62368-1 / UL62368-1, LVD IEC60601-1, EAC TP TC 020/2011 approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC
- ISOLATION CAPACITANCE(Typ.)
- 10pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55032, BS EN/EN55011 Class A/B with external components Radiated BS EN/EN55032, BS EN/EN55011 Class A/B with external components - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 2, ±4KV contact, Level 3, ±8KV air Radiated Susceptibility BS EN/EN61000-4-3 Level 2, 3V/m EFT/Bursts BS EN/EN61000-4-4 Level 1, 0.5KV / 5KHz Surge BS EN/EN61000-4-5 Level 1, 0.5KV Line to Ground Conducted BS EN/EN61000-4-6 Level 2, 3V(e.m.f.) Magnetic Field BS EN/EN61000-4-8 Level 1, 1A/m
OTHERS
- MTBF
- 15000Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 19.7*7.1*11.5mm (0.775*0.279*0.452 inch)
- CASE MATERIAL
- Non-Conductive black plastic (UL 94V-0 rated)
- PACKING
- 2.8g ; 24pcs/per tube, 2160pcs/90 tube/per carton
Test conditions and notes6
- 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load.
- 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
- Telecom / datacom system
- Wireless network
- Industrial control facility
Downloads
Datasheet1
FAQ
What is the Mean Well MDS01M-12N designed for?
The MDS01M-12N is a 1W SIP-package DC-DC converter intended for compact embedded and industrial electronics that need a regulated 12V DC output. Its reinforced 6KVDC input-to-output isolation and medical safety approvals also make it relevant for equipment where isolation and low leakage current matter.
What are the key electrical specifications of the MDS01M-12N?
This model provides a single 12V DC output with a 1W power rating and a rated output current of 15mA. The product family supports several nominal input ranges, and the 12V-input version is specified for 10.8 to 13.2V DC input.
Is the MDS01M-12N suitable for medical or safety-sensitive equipment?
Yes. The product description states compliance with IEC/UL 62368-1 and LVD IEC60601-1, and it is medical safety approved to 1xMOPP/2xMOOP according to IEC60601-1. It also specifies low patient leakage current below 2uA, which is important in medical-related applications.
How does the MDS01M-12N perform in harsh temperature conditions?
The MDS01M-12N is rated for operation from -40C to +95C. That wide operating range makes it a practical option for equipment exposed to demanding industrial or enclosed-environment temperatures.
What protection and isolation features does this converter include?
This model includes continuous short-circuit protection and overload protection that shuts down the output when rated power reaches 102% or more, then automatically recovers. It also provides 6KVDC input-to-output isolation with reinforced insulation for applications that require strong electrical separation.
Why would a buyer choose the MDS01M-12N over a larger power converter?
The MDS01M-12N is optimized for low-power applications where compact size, isolation, and compliance matter more than higher output wattage. Its small 19.7 x 7.1 x 11.5 mm SIP7 form factor, 1W output, and medical-oriented safety credentials differentiate it from larger general-purpose DC-DC converters.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other MDS01 Models
11 other modelsPlease log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.