Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
MDD02N-12 is the ±12V output model of this series: rated power 2W, typical efficiency 82%.
MDS02 and MDD02 series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP 7 package. It features international standard pins, a high efficiency up to 8 5 %, wide working temperature range -40~+ 85℃, 6 KVDC or 4 . 2KVAC I/P-O/P hight isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 5V/12V/ 15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current.
Features
- SIP7 package with international standard pinout
- Operating temperature range -40 ~ +85℃
- Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
- Low patient leakage current <2μA
- Protection: Short circuit(3 second max.)
- 6KVDC or 4.2VAC hight I/O isolation (Reinforced isolation)
- Low cost
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 24V (21.6 ~ 26.4V)
- INPUT CURRENT NO LOAD
- 11mA
- INPUT CURRENT FULL LOAD
- 103mA
- VOLTAGE RANGE
- 21.6 ~ 26.4Vdc
- SURGE VOLTAGE (100ms max.)
- 5Vin models : 9Vdc 12Vin models : 18Vdc 24Vin models : 30Vdc
- FILTER
- Internal capacitor
- PROTECTION
- Fuse recommended. 500mA Slow-Blow Type for all models
OUTPUT
- OUTPUT VOLTAGE
- ±12V
- OUTPUT CURRENT
- ±9 ~ 83mA
- EFFICIENCY (TYP.)
- 82%
- CAPACITOR LOAD (MAX.)* For each output
- *220μF
- VOLTAGE ACCURACY
- ±5.0%
- RATED POWER
- 2W
- RIPPLE & NOISENote.2
- 75mVp-p
- LINE REGULATIONNote.3
- 1.2% for 1% input variation
- LOAD REGULATIONNote.4
- ±10%
- SWITCHING FREQUENCY (Typ.)
- 100KHz
PROTECTION
- SHORT CIRCUIT
- 3 second max.
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +85℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.02% / ℃ (0 ~ 85℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC ( Note.6)
- SAFETY STANDARDS
- UL60601-1, EAC TP TC 020/2011 approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 5pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011(CISPR11) Class B Radiated BS EN/EN55011(CISPR11) Class B - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 2, ±8KV contact
OTHERS
- MTBF
- 3500Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 19.5*9.8*12.5mm (0.77*0.386*0.492 inch)
- CASE MATERIAL
- Non-Conductive black plastic (UL 94V-0 rated)
- PACKING
- 4.2g ; 25pcs/per tube, 3000pcs/120 tube/per carton
Test conditions and notes6
- 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load.
- 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
Downloads
Certificates3
Test reports1
FAQ
Can I use the MDD02N-12 in a 24V industrial control cabinet that needs isolated +/-12V rails for analog or interface circuits?
It can fit 24V-class control systems that stay within the converter's 21.6~26.4Vdc input range and need an unregulated +/-12V output. The available output is 2W total with up to 83mA per output, so it is suited to low-power analog, sensing, or interface sections rather than heavier loads.
Is the MDD02N-12 appropriate for embedded control boards that need a small +/-12V auxiliary supply?
This model is sized for compact embedded designs because it provides 2W in a 19.5 x 9.8 x 12.5 mm package. Its unregulated +/-12V output and 0.0042 kg weight make it a fit for low-power onboard auxiliary rails where space is limited.
Can this converter be used for low-power instrumentation or signal-conditioning circuits in automation equipment?
Yes, it is aligned with low-power instrumentation use where a +/-12V rail is needed and the load stays within 2W and 83mA per output; note the output is unregulated (load regulation ±10%). The -40°C to +85°C operating range also supports use in demanding industrial temperature conditions.
Is the MDD02N-12 suitable for outdoor equipment or washdown environments?
The listed specs do not include an IP rating, so this converter should not be treated as outdoor- or washdown-rated by itself. Its environmental data is the -40°C to +85°C operating range; enclosure-level protection would need to come from the host system.
How does the MDD02N-12 compare with the other MDD02N models?
Within the MDD02N series, all models are 2W parts with 21.6~26.4V input and different outputs: MDD02N-05 is +/-5V, +/-20~200mA; MDD02N-09 is +/-9V, +/-12~111mA; MDD02N-12 is +/-12V, +/-9~83mA; and MDD02N-15 is +/-15V, +/-7~67mA. The main selector in this group is the output voltage and output current for the required rail.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other MDD02 Models
20 other modelsPlease log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.