MDS01L-12 is the 12V output model of this series: rated power 1W, typical efficiency 77%.
MDS01 and MDD01 series are 1W isolated and unregulated module type medical grade DC-DC converter with SIP7 package. It features international standard pins, a high efficiency up to 83%, wide working temperature range -40~+85℃, 6KVDC or 4.2KVAC I/P-O/P high isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 3.3V/ 5V/ 12 V/15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems , ultra low leakage current.
Features
- SIP7 package with international standard pinout
- Operating temperature range -40 ~ +85℃
- Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
- Low patient leakage current <2μA
- Protection: Short circuit(3 second max.)
- 6KVDC or 4.2KVAC hight I/O isolation (Reinforced isolation)
- Low cost
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 5V (4.5 ~ 5.5V)
- VOLTAGE RANGE
- 4.5 ~ 5.5Vdc
- INPUT CURRENT (NO LOAD)
- 40mA
- INPUT CURRENT (FULL LOAD)
- 260mA
- SURGE VOLTAGE (100ms max.)
- 5Vin models : 9Vdc 12Vin models : 18Vdc 24Vin models : 30Vdc
- FILTER
- Internal capacitor
- PROTECTION
- Fuse recommended. 500mA Slow-Blow Type for all models
OUTPUT
- OUTPUT VOLTAGE
- 12V
- OUTPUT CURRENT
- 9 ~ 84mA
- EFFICIENCY (TYP.)
- 77%
- CAPACITOR LOAD (MAX.)* For each output
- 470μF
- VOLTAGE ACCURACY
- ±5.0%
- RATED POWER
- 1W
- RIPPLE & NOISENote.2
- 75mVp-p
- LINE REGULATIONNote.3
- 1.2% for 1% input variation
- LOAD REGULATIONNote.4
- ±10%
- SWITCHING FREQUENCY (Typ.)
- 100KHz
PROTECTION
- SHORT CIRCUIT
- 3 second max.
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +85℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.02% / ℃ (0 ~ 85℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC
- SAFETY STANDARDS
- UL60601-1, EAC TP TC 020/2011 approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 5pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011(CISPR11) Class B Radiated BS EN/EN55011(CISPR11) Class B - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 2, ±8KV contact
OTHERS
- MTBF
- 3500Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 19.5*9.8*12.5mm (0.77*0.386*0.492 inch)
- CASE MATERIAL
- Non-Conductive black plastic (UL 94V-0 rated)
- PACKING
- 4.2g ; 25pcs/per tube, 3000pcs/120 tube/per carton
Test conditions and notes6
- 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load.
- 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
Downloads
Certificates5
Test reports1
FAQ
Can I use the MDS01L-12 for a small 12V signal-conditioning or monitoring circuit?
Yes, it is suited to low-power 12V electronics that fit within its 1W output budget. The converter provides a fixed, unregulated 12V output up to 84mA, which aligns with compact monitoring, interface, and signal-conditioning sections rather than heavier 12V loads.
Is this converter appropriate for embedded 5V-to-12V control-board designs?
It is a fit for embedded designs that need a low-power 12V rail from a 4.5~5.5VDC input. The unit is rated at 1W and delivers 12V at up to 84mA, so it works best for light auxiliary circuits on compact boards.
Can the MDS01L-12 be used in industrial or automation equipment with tight space limits?
Yes, its 19.5 x 9.8 x 12.5 mm body and 0.0042 kg weight make it practical for dense control electronics. Its 1W rating and fixed 12V output are better matched to low-power internal electronics than to field devices with higher current demand.
Is the MDS01L-12 suitable for outdoor or harsh-environment use?
The operating temperature range is -40 C to +85 C (with load derating at higher temperatures, see the derating curve), so it covers a wide thermal range for equipment design. No IP rating is listed, so the product data does not indicate weather exposure or washdown suitability on its own.
How does the MDS01L-12 compare with other MDS01L series options?
Within the MDS01L family, all models are 1W parts, but output options vary: 3.3V on MDS01L-03, 5V on MDS01L-05, 12V on MDS01L-12, and 15V on MDS01L-15. Their maximum output currents are 303mA, 200mA, 84mA, and 67mA respectively.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other MDS01 Models
12 other modelsPlease log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.