MDS02L-12 is the 12V output model of this series: rated power 2W, typical efficiency 80%.
MDS02 and MDD02 series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP 7 package. It features international standard pins, a high efficiency up to 8 5 %, wide working temperature range -40~+ 85℃, 6 KVDC or 4 . 2KVAC I/P-O/P hight isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 5V/12V/ 15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current.
Features
- SIP7 package with international standard pinout
- Operating temperature range -40 ~ +85℃
- Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
- Low patient leakage current <2μA
- Protection: Short circuit(3 second max.)
- 6KVDC or 4.2VAC hight I/O isolation (Reinforced isolation)
- Low cost
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 5V (4.5 ~ 5.5V)
- INPUT CURRENT NO LOAD
- 47mA
- INPUT CURRENT FULL LOAD
- 510mA
- VOLTAGE RANGE
- 4.5 ~ 5.5Vdc
- SURGE VOLTAGE (100ms max.)
- 5Vin models : 9Vdc 12Vin models : 18Vdc 24Vin models : 30Vdc
- FILTER
- Internal capacitor
- PROTECTION
- Fuse recommended. 500mA Slow-Blow Type for all models
OUTPUT
- OUTPUT VOLTAGE
- 12V
- OUTPUT CURRENT
- 17 ~ 167mA
- EFFICIENCY (TYP.)
- 80%
- CAPACITOR LOAD (MAX.)* For each output
- 470μF
- VOLTAGE ACCURACY
- ±5.0%
- RATED POWER
- 2W
- RIPPLE & NOISENote.2
- 75mVp-p
- LINE REGULATIONNote.3
- 1.2% for 1% input variation
- LOAD REGULATIONNote.4
- ±10%
- SWITCHING FREQUENCY (Typ.)
- 100KHz
PROTECTION
- SHORT CIRCUIT
- 3 second max.
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +85℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.02% / ℃ (0 ~ 85℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC ( Note.6)
- SAFETY STANDARDS
- UL60601-1, EAC TP TC 020/2011 approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 5pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011(CISPR11) Class B Radiated BS EN/EN55011(CISPR11) Class B - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 2, ±8KV contact
OTHERS
- MTBF
- 3500Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 19.5*9.8*12.5mm (0.77*0.386*0.492 inch)
- CASE MATERIAL
- Non-Conductive black plastic (UL 94V-0 rated)
- PACKING
- 4.2g ; 25pcs/per tube, 3000pcs/120 tube/per carton
Test conditions and notes6
- 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load.
- 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
Downloads
FAQ
Can I use the MDS02L-12 for low-power 12V electronics in embedded hardware?
This model is rated for a 4.5~5.5Vdc input and provides an unregulated 12V output at 2W. That makes it appropriate for low-power embedded hardware sections that require a 12V rail from that input range.
Is the MDS02L-12 suitable for 12V auxiliary circuits in industrial control equipment?
Yes, if the industrial control equipment has a 4.5~5.5Vdc source available and the auxiliary circuit fits within the 2W rating. It is intended for 12V support electronics rather than larger 12V power loads.
Can this converter be used for 12V automation or instrumentation modules?
It can be used where an automation or instrumentation module needs a 12V output from a 4.5~5.5Vdc input. Its compact size of 19.5 × 9.8 × 12.5 mm is useful where module space is tight.
Is the MDS02L-12 appropriate for harsh indoor environments?
The operating temperature is specified from -40°C to +85°C, which supports a wide range of indoor equipment conditions. Because no IP rating is listed, it is better suited to protected enclosures than exposed wet or outdoor environments.
What is the difference between the MDS02L-12 and the other MDS02L models?
The MDS02L family shown on the sheet is entirely 2W with 4.5~5.5V input. The other versions are MDS02L-05 at 5V and 40~400mA and MDS02L-15 at 15V and 13~133mA, while this model is 12V at 17~167mA.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other MDS02 Models
12 other modelsPlease log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.