Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
MDS02M-12 is the 12V output model of this series: rated power 2W, typical efficiency 83%.
MDS02 and MDD02 series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP 7 package. It features international standard pins, a high efficiency up to 8 5 %, wide working temperature range -40~+ 85℃, 6 KVDC or 4 . 2KVAC I/P-O/P hight isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 5V/12V/ 15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current.
Features
- SIP7 package with international standard pinout
- Operating temperature range -40 ~ +85℃
- Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
- Low patient leakage current <2μA
- Protection: Short circuit(3 second max.)
- 6KVDC or 4.2VAC hight I/O isolation (Reinforced isolation)
- Low cost
- 3 years warranty
Specifications
INPUT
- INPUT VOLTAGE (RANGE)
- Normal 12V (10.8 ~ 13.2V)
- INPUT CURRENT NO LOAD
- 16mA
- INPUT CURRENT FULL LOAD
- 205mA
- VOLTAGE RANGE
- 10.8 ~ 13.2Vdc
- SURGE VOLTAGE (100ms max.)
- 5Vin models : 9Vdc 12Vin models : 18Vdc 24Vin models : 30Vdc
- FILTER
- Internal capacitor
- PROTECTION
- Fuse recommended. 500mA Slow-Blow Type for all models
OUTPUT
- OUTPUT VOLTAGE
- 12V
- OUTPUT CURRENT
- 17 ~ 167mA
- EFFICIENCY (TYP.)
- 83%
- CAPACITOR LOAD (MAX.)* For each output
- 470μF
- VOLTAGE ACCURACY
- ±5.0%
- RATED POWER
- 2W
- RIPPLE & NOISENote.2
- 75mVp-p
- LINE REGULATIONNote.3
- 1.2% for 1% input variation
- LOAD REGULATIONNote.4
- ±10%
- SWITCHING FREQUENCY (Typ.)
- 100KHz
PROTECTION
- SHORT CIRCUIT
- 3 second max.
ENVIRONMENT
- COOLING
- Free-air convection
- WORKING TEMP.
- -40 ~ +85℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -55 ~ +125℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- 0.02% / ℃ (0 ~ 85℃)
- SOLDERING TEMPERATURE
- 1.5mm from case of 1 ~ 3sec./260℃ max.
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC ( Note.6)
- SAFETY STANDARDS
- UL60601-1, EAC TP TC 020/2011 approved
- WITHSTAND VOLTAGE
- I/P-O/P:6KVDC or 4.2KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- ISOLATION LEVEL
- Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC, 50/60Hz
- ISOLATION CAPACITANCE (Typ.)
- 5pF
- EMC EMISSION
-
Parameter Standard Test Level Conducted BS EN/EN55011(CISPR11) Class B Radiated BS EN/EN55011(CISPR11) Class B - EMC IMMUNITY
-
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 2, ±8KV contact
OTHERS
- MTBF
- 3500Khrs MIL-HDBK-217F(25℃)
- DIMENSION (L*W*H)
- 19.5*9.8*12.5mm (0.77*0.386*0.492 inch)
- CASE MATERIAL
- Non-Conductive black plastic (UL 94V-0 rated)
- PACKING
- 4.2g ; 25pcs/per tube, 3000pcs/120 tube/per carton
Test conditions and notes6
- 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
- 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
- 3.Line regulation is measured from low line to high line at rated load.
- 4.Load regulation is measured from 10% to 100% rated load.
- 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
- 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)
Technical drawings
Typical applications
- Medical devices
- Medical oxygen monitor
- CT scanning
- Medical carts
- Oral care equipment
Downloads
FAQ
Can I use the MDS02M-12 to supply a 12V analog or interface rail from a 12V-class input source?
Yes, this unit provides an unregulated 12V output from a 10.8~13.2V input range. Its 2W rating makes it suitable for low-power analog, sensing, or interface circuits rather than heavier 12V loads.
Is the MDS02M-12 suitable for small automation modules that need an isolated-style board-level 12V rail?
It is sized for compact board-level power conversion, with dimensions of 19.5 × 9.8 × 12.5 mm and a weight of 0.0042 kg. The 2W output suits low-power sections inside industrial or embedded control assemblies.
Can this converter be used in embedded test, measurement, or industrial sensing electronics that require 12V regulation?
It can fit low-power 12V rails in embedded measurement or sensing electronics when the available input is 10.8~13.2V, but note the output is unregulated (load regulation ±10%). The product is rated for 2W, so it is best aligned with auxiliary electronics and signal-level subsystems.
How does the operating range affect where the MDS02M-12 can be used?
The specified operating temperature is -40°C to +85°C, which supports use in a wide range of industrial temperature conditions. No IP rating appears in the listed specs, so the information provided supports protected-enclosure use rather than direct weather exposure.
How is the MDS02M-12 positioned within the other MDS02M output variants?
The MDS02M family covers 5V, 12V, and 15V outputs at 2W with 10.8~13.2V input. The sibling models are MDS02M-05 at 5V and 40~400mA and MDS02M-15 at 15V and 13~133mA, while the MDS02M-12 delivers 12V at 17~167mA.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.