MDS02N-12N is the 12V output model of this series: rated power 2W, typical efficiency 84%.

MDS02-N and MDD02-N series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP7 package. It features international standard pins, a high efficiency up to 86%, wide working temperature range -40~+95℃, 6KVDC I/P-O/P high isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 3.3V/5V/12V/15V for single output and ±5V /±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current or industrial control, telecommunication field, distributed power architecture, and so on.

Features

  • SIP7 package with international standard pinout
  • Operating temperature range -40 ~ +95℃
  • Comply with IEC/UL62368-1 and LVD IEC60601-1
  • Medical safety approved (1xMOPP/2xMOOP) according to IEC60601-1
  • Low patient leakage current <2μA
  • Protections: Short circuit (Continuous)
  • 6KVDC hight I/O isolation (Reinforced insulation)
  • Low cost
  • 3 years warranty

Specifications

MDS02N-12NSpecifications from the official datasheet (2024-08-12 edition)

MODEL SELECTION TABLE

INPUT VOLTAGE (RANGE)
Normal 24V (21.6 ~ 26.4V)
INPUT CURRENT NO LOAD
12mA
INPUT CURRENT FULL LOAD
100mA
OUTPUT VOLTAGE
12V
OUTPUT CURRENT
17 ~ 167mA
EFFICIENCY (TYP.)
84%
CAPACITOR LOAD (MAX.)* For each output
330μF

INPUT

VOLTAGE RANGE
21.6 ~ 26.4Vdc
FILTER
Internal capacitor
PROTECTION
Fuse recommended. 500mA Slow-Blow Type for all models

OUTPUT

VOLTAGE ACCURACY
±5.0%
RATED POWER
2W
RIPPLE & NOISENote.2
100mVp-p
LINE REGULATIONNote.3
1.2% for 1% input variation
LOAD REGULATIONNote.4
±15% for 3.3Vo, 5Vo ; ±10% for others
SWITCHING FREQUENCY (Typ.)
20KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +95℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+105℃
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.02% / ℃ (0 ~ 90℃)
SOLDERING TEMPERATURE
1.5mm from case of 1 ~ 3sec./260℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.6)

SAFETY STANDARDS
IEC62368-1 / UL62368-1, LVD IEC60601-1, EAC TP TC 020/2011 approved
WITHSTAND VOLTAGE
I/P-O/P:6KVDC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION LEVEL
Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC
ISOLATION CAPACITANCE (Typ.)
25pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55032, BS EN/EN55011Class A/B with external components
RadiatedBS EN/EN55032, BS EN/EN55011Class A/B with external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 2, ±4KV contact, Level 3, ±8KV air
Radiated SusceptibilityBS EN/EN61000-4-3Level 2, 3V/m
EFT/BurstsBS EN/EN61000-4-4Level 1, 0.5KV / 5KHz
SurgeBS EN/EN61000-4-5Level 1, 0.5KV Line to Ground
ConductedBS EN/EN61000-4-6Level 2, 3V(e.m.f.)
Magnetic FieldBS EN/EN61000-4-8Level 1, 1A/m

OTHERS

MTBF
12900Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
19.7*7.1*11.5mm (0.775*0.279*0.452 inch)
CASE MATERIAL
Non-Conductive black plastic (UL 94V-0 rated)
PACKING
2.8g ; 24pcs/per tube, 2160pcs/90 tube/per carton
Test conditions and notes6
  1. 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 10% to 100% rated load.
  5. 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
  6. 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS02N-12N Mechanical Specification
Mechanical Specification
MDS02N-12N EMC Suggestion
EMC Suggestion
MDS02N-12N Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
  • Telecom / datacom system
  • Wireless network
  • Industrial control facility

Downloads

::specification::

FAQ

What is the Mean Well MDS02N-12N used for?

The MDS02N-12N is a 2W SIP-package, medical grade unregulated DC-DC converter intended for compact embedded power applications that need a 12V output. Its reinforced 6KVDC input-to-output isolation and medical safety approval make it relevant for industrial and medical-related equipment where isolation and safety matter.

What are the key electrical specifications of the MDS02N-12N?

This model is rated for 2W output power and provides a 12V DC output with an output current range of 17 to 167mA. It is the 24V nominal input version and operates over a 21.6 to 26.4V DC input range.

Is the MDS02N-12N suitable for medical or safety-sensitive equipment?

Yes, this series is medical safety approved to IEC60601-1 with 1xMOPP and 2xMOOP. It also lists a patient leakage current of 2μA max and approval to IEC62368-1/UL62368-1 and LVD IEC60601-1, which supports use in applications with stricter safety requirements.

How does the MDS02N-12N perform in harsh temperature conditions?

The MDS02N-12N has an operating temperature range of -40°C to +95°C, with load derating at higher temperatures per the derating curve. That wide range makes it a practical option for equipment exposed to cold starts, elevated internal enclosure temperatures, or other demanding operating environments.

What protection features does the MDS02N-12N include?

This converter includes continuous short-circuit protection with automatic recovery. Mean Well also recommends an external 500mA slow-blow fuse on the input.

Why would a buyer choose the MDS02N-12N over a non-isolated compact converter?

A main differentiator is its 6KVDC input-to-output isolation with reinforced insulation in a compact SIP package. Buyers comparing isolated versus non-isolated options may choose this model when safety approvals, isolation performance, and small size are more important than using a basic low-cost converter.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":9596818719019, "variants" :[{ "context":"default", "variant_id": 49368192483627, "sku":"MDS02N-12N", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS02N-12N SIP Package DC-DC Converter- 2W - 12V Out 167mA
Mean Well

Mean Well MDS02N-12N SIP Package DC-DC Converter- 2W - 12V Out 167mA

MDS02N-12N is the 12V output model of this series: rated power 2W, typical efficiency 84%.

MDS02-N and MDD02-N series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP7 package. It features international standard pins, a high efficiency up to 86%, wide working temperature range -40~+95℃, 6KVDC I/P-O/P high isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 3.3V/5V/12V/15V for single output and ±5V /±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current or industrial control, telecommunication field, distributed power architecture, and so on.

Features

  • SIP7 package with international standard pinout
  • Operating temperature range -40 ~ +95℃
  • Comply with IEC/UL62368-1 and LVD IEC60601-1
  • Medical safety approved (1xMOPP/2xMOOP) according to IEC60601-1
  • Low patient leakage current <2μA
  • Protections: Short circuit (Continuous)
  • 6KVDC hight I/O isolation (Reinforced insulation)
  • Low cost
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
  • Telecom / datacom system
  • Wireless network
  • Industrial control facility