Overview

MDS15A-15 is the 15V output model of this series: rated power 15W, typical efficiency 87%.

MDS15 series is 15W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 88%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Specifications

MDS15A-15Specifications from the official datasheet (2024-04-30 edition)

MODEL SELECTION TABLE

INPUT VOLTAGE (RANGE)
Normal 12V (9 ~ 18V)
INPUT CURRENT NO LOAD
10mA
INPUT CURRENT FULL LOAD
1450mA
OUTPUT VOLTAGE
15V
OUTPUT CURRENT
1000mA
EFFICIENCY (Typ.)
87%
CAPACITOR LOAD (MAX.)
720μF

INPUT

VOLTAGE RANGE
9~18Vdc
SURGE VOLTAGE (100ms max.)
12Vin models : 25Vdc, 24Vin models : 50Vdc, 48Vin models : 100Vdc
FILTER
Pi type
PROTECTION
Fuse recommended. 12Vin models: 4A delay time Type, 24Vin models: 2A delay time Type, 48Vin models: 1A delay time Type

OUTPUT

VOLTAGE ACCURACY
±1%
RATED POWER
15W
RIPPLE & NOISENote.2
60mVp-p
LINE REGULATIONNote.3
±0.5%
LOAD REGULATIONNote.4
±0.5%
SWITCHING FREQUENCY (Typ.)
12Vin/24Vin:250KHz, 48Vin:300KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery
OVERLOAD
110 ~ 185% rated output power Protection type : Recovers automatically after fault condition is removed
UNDER VOLTAGE LOCKOUT
12Vin: 7.5Vdc, 24Vin: 15Vdc, 48Vin: 33Vdc

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +90℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+110℃ max.
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.03% / ℃ (0 ~ 70℃)
SOLDERING TEMPERATURE
1.5mm from case of 1 ~ 3sec./260℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.6)

SAFETY STANDARDS
IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
WITHSTAND VOLTAGE
I/P-O/P 4KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION CAPACITANCE (Typ.)
100pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011Class A without external components
RadiatedBS EN/EN55011Class A without external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 1, ±15KV air, ±8KV contact
Radiated SusceptibilityBS EN/EN61000-4-3Level 1, 10V/m
EFT/BurstsBS EN/EN61000-4-4Level 1, ±2KV
SurgeBS EN/EN61000-4-5Level 1, ±1KV Line-Line, ±2KV Line-GND
ConductedBS EN/EN61000-4-6Level 1, 10Vrms(e.m.f.)
Magnetic FieldBS EN/EN61000-4-8Level 2, 10A/m

OTHERS

MTBF
1060Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
50.8*25.4*12mm (2*1*0.47 inch)
CASE MATERIAL
UL94V-0 plastic case
PACKING
30g; 18pcs/per tube, 432pcs/24 tube max./carton
Test conditions and notes6
  1. 1.All parameters are specified at normal input(A:12Vdc, B:24Vdc, C:48Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 0% to 100% rated load.
  5. 5.2xMOPP base on a working voltage of 250Vac between Primary to Secondary.
  6. 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS15A-15 Pin Configuration
Pin Configuration
MDS15A-15 Derating Curve
Derating Curve
MDS15A-15 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment

Downloads

FAQ

Can I use the MDS15A-15 for 15V industrial instrumentation or analog control circuits?

Yes, it is designed as a constant-voltage 15V output unit rated for 15W and 1A. That makes it suitable for low-power 15V industrial or control electronics when the input source is within 9-18Vdc.

Is the MDS15A-15 a fit for embedded systems that need a regulated 15V rail?

It can be used where the load specifically requires 15V constant-voltage output and stays within 1A. The input side also needs to match the listed 9-18Vdc range.

Can this model support low-power 15V automation modules inside equipment enclosures?

For compact automation modules, the main fit points are its 15V output, 15W rating, and 1A rated output current. It is best matched to low-power electronics rather than loads that need more than 15W.

Will the MDS15A-15 handle cold or high-heat operating environments?

The specified operating range is -40C to +90C (load derating applies at higher ambient temperatures per the derating curve), so the temperature rating is wide for industrial environments. The specs do not include an IP rating, so temperature capability is stated, but ingress protection is not.

What are the differences between the MDS15A-15 and the other MDS15A versions?

The MDS15A family shares a 15W rating and 9-18Vdc input range. Output options are 5V at 3A on MDS15A-05, 12V at 1.25A on MDS15A-12, 15V at 1A on MDS15A-15, and 24V at 625mA on MDS15A-24.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8442611761451, "variants" :[{ "context":"default", "variant_id": 45711618081067, "sku":"MDS15A-15", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS15A-15 DC-DC Converter - 15W - 9~18V in 15V out 1A
Mean Well

Mean Well MDS15A-15 DC-DC Converter - 15W - 9~18V in 15V out 1A

MDS15A-15 is the 15V output model of this series: rated power 15W, typical efficiency 87%.

MDS15 series is 15W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 88%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
View product