💡 New range: Boke DALI-2 LED drivers | 10–240 W | 5-year warranty Explore the range

Over 15 years experience of providing support of MEAN WELL Power Supplies

Image shown is a representation only. Exact specifications should be obtained from the product datasheet.

Mean Well MDS15B-15 DC-DC Converter - 15W - 18~36V in 15V out 1A

SKU MDS15B-15
Log in to see price Trade pricing & online ordering for account holders
Available to order
|

Overview

MDS15B-15 is the 15V output model of this series: rated power 15W, typical efficiency 86%.

MDS15 series is 15W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 88%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Specifications

MDS15B-15Specifications from the official datasheet (2024-04-30 edition)

MODEL SELECTION TABLE

INPUT VOLTAGE (RANGE)
Normal 24V (18 ~ 36V)
INPUT CURRENT NO LOAD
7mA
INPUT CURRENT FULL LOAD
700mA
OUTPUT VOLTAGE
15V
OUTPUT CURRENT
1000mA
EFFICIENCY (Typ.)
86%
CAPACITOR LOAD (MAX.)
720μF

INPUT

VOLTAGE RANGE
18~36Vdc
SURGE VOLTAGE (100ms max.)
12Vin models : 25Vdc, 24Vin models : 50Vdc, 48Vin models : 100Vdc
FILTER
Pi type
PROTECTION
Fuse recommended. 12Vin models: 4A delay time Type, 24Vin models: 2A delay time Type, 48Vin models: 1A delay time Type

OUTPUT

VOLTAGE ACCURACY
±1%
RATED POWER
15W
RIPPLE & NOISENote.2
60mVp-p
LINE REGULATIONNote.3
±0.5%
LOAD REGULATIONNote.4
±0.5%
SWITCHING FREQUENCY (Typ.)
12Vin/24Vin:250KHz, 48Vin:300KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery
OVERLOAD
110 ~ 185% rated output power Protection type : Recovers automatically after fault condition is removed
UNDER VOLTAGE LOCKOUT
12Vin: 7.5Vdc, 24Vin: 15Vdc, 48Vin: 33Vdc

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +90℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+110℃ max.
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.03% / ℃ (0 ~ 70℃)
SOLDERING TEMPERATURE
1.5mm from case of 1 ~ 3sec./260℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.6)

SAFETY STANDARDS
IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
WITHSTAND VOLTAGE
I/P-O/P 4KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION CAPACITANCE (Typ.)
100pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011Class A without external components
RadiatedBS EN/EN55011Class A without external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 1, ±15KV air, ±8KV contact
Radiated SusceptibilityBS EN/EN61000-4-3Level 1, 10V/m
EFT/BurstsBS EN/EN61000-4-4Level 1, ±2KV
SurgeBS EN/EN61000-4-5Level 1, ±1KV Line-Line, ±2KV Line-GND
ConductedBS EN/EN61000-4-6Level 1, 10Vrms(e.m.f.)
Magnetic FieldBS EN/EN61000-4-8Level 2, 10A/m

OTHERS

MTBF
1060Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
50.8*25.4*12mm (2*1*0.47 inch)
CASE MATERIAL
UL94V-0 plastic case
PACKING
30g; 18pcs/per tube, 432pcs/24 tube max./carton
Test conditions and notes6
  1. 1.All parameters are specified at normal input(A:12Vdc, B:24Vdc, C:48Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 0% to 100% rated load.
  5. 5.2xMOPP base on a working voltage of 250Vac between Primary to Secondary.
  6. 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS15B-15 Pin Configuration
Pin Configuration
MDS15B-15 Derating Curve
Derating Curve
MDS15B-15 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment

Downloads

FAQ

Can I use the MDS15B-15 to provide a 15V rail inside industrial control equipment?

Yes, this model is a 15W constant-voltage DC-DC converter with a 15V output at 1A. That makes it suitable for low-power 15V auxiliary rails in industrial control, embedded, or automation equipment that is fed from an 18-36Vdc input range.

Is the MDS15B-15 a fit for 24V-class vehicle or battery-based systems that need 15V output?

It can fit systems where the available source stays within the listed 18-36Vdc input range and the load needs 15V constant voltage. With 15W total output power and 1A output current, it is aimed at lighter-duty embedded or control loads rather than larger field devices.

Can I use this converter for instrumentation or analog interface boards that require 15V?

Yes, the 15V constant-voltage output is appropriate for low-power instrumentation, analog interface, and signal-conditioning circuits when the total demand stays within 15W. The compact 50.8 x 25.4 x 12.0 mm size also makes it easier to place in dense control assemblies.

Is the MDS15B-15 suitable for PLC cabinet or embedded enclosure environments?

It is rated for operation from -40 C to +90 C (load derating applies at higher ambient temperatures per the derating curve), which supports use in a wide range of industrial indoor environments. However, no IP rating is listed in the provided specs, so the specifications do not establish outdoor exposure resistance and it is better treated as a converter for protected enclosures.

How does the MDS15B-15 compare with the other MDS15B models?

All listed MDS15B models are 15W converters in the same family. The sibling outputs are MDS15B-05 at 5V and 3A, MDS15B-12 at 12V and 1.25A, MDS15B-15 at 15V and 1A, and MDS15B-24 at 24V and 625mA.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8442612187435, "variants" :[{ "context":"default", "variant_id": 45711618801963, "sku":"MDS15B-15", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS15B-15 DC-DC Converter - 15W - 18~36V in 15V out 1A
Mean Well

Mean Well MDS15B-15 DC-DC Converter - 15W - 18~36V in 15V out 1A

MDS15B-15 is the 15V output model of this series: rated power 15W, typical efficiency 86%.

MDS15 series is 15W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 88%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
View product