MDS02L-05 is the 5V output model of this series: rated power 2W, typical efficiency 77%.

MDS02 and MDD02 series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP 7 package. It features international standard pins, a high efficiency up to 8 5 %, wide working temperature range -40~+ 85℃, 6 KVDC or 4 . 2KVAC I/P-O/P hight isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 5V/12V/ 15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current.

Features

  • SIP7 package with international standard pinout
  • Operating temperature range -40 ~ +85℃
  • Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
  • Low patient leakage current <2μA
  • Protection: Short circuit(3 second max.)
  • 6KVDC or 4.2VAC hight I/O isolation (Reinforced isolation)
  • Low cost
  • 3 years warranty

Specifications

MDS02L-05Specifications from the official datasheet (2022-05-24 edition)

INPUT

INPUT VOLTAGE (RANGE)
Normal 5V (4.5 ~ 5.5V)
INPUT CURRENT NO LOAD
35mA
INPUT CURRENT FULL LOAD
510mA
VOLTAGE RANGE
4.5 ~ 5.5Vdc
SURGE VOLTAGE (100ms max.)
5Vin models : 9Vdc 12Vin models : 18Vdc 24Vin models : 30Vdc
FILTER
Internal capacitor
PROTECTION
Fuse recommended. 500mA Slow-Blow Type for all models

OUTPUT

OUTPUT VOLTAGE
5V
OUTPUT CURRENT
40 ~ 400mA
EFFICIENCY (TYP.)
77%
CAPACITOR LOAD (MAX.)* For each output
1000μF
VOLTAGE ACCURACY
±5.0%
RATED POWER
2W
RIPPLE & NOISENote.2
75mVp-p
LINE REGULATIONNote.3
1.2% for 1% input variation
LOAD REGULATIONNote.4
±10%
SWITCHING FREQUENCY (Typ.)
100KHz

PROTECTION

SHORT CIRCUIT
3 second max.

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +85℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.02% / ℃ (0 ~ 85℃)
SOLDERING TEMPERATURE
1.5mm from case of 1 ~ 3sec./260℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.6)

SAFETY STANDARDS
UL60601-1, EAC TP TC 020/2011 approved
WITHSTAND VOLTAGE
I/P-O/P:6KVDC or 4.2KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION LEVEL
Primary-secondary: 1xMOPP / 2xMOOP when system input voltage is with 250VAC, 50/60Hz
ISOLATION CAPACITANCE (Typ.)
5pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011(CISPR11)Class B
RadiatedBS EN/EN55011(CISPR11)Class B
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 2, ±8KV contact

OTHERS

MTBF
3500Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
19.5*9.8*12.5mm (0.77*0.386*0.492 inch)
CASE MATERIAL
Non-Conductive black plastic (UL 94V-0 rated)
PACKING
4.2g ; 25pcs/per tube, 3000pcs/120 tube/per carton
Test conditions and notes6
  1. 1.All parameters are specified at normal input(L:5Vdc, M:12Vdc, N:24Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 10% to 100% rated load.
  5. 5.Patient leakage current(2μA max.) and reinforced isolation is based on a 250VAC, 50/60Hz system input voltage.
  6. 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS02L-05 Mechanical Specification
Mechanical Specification
MDS02L-05 EMC Suggestion
EMC Suggestion
MDS02L-05 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment

Downloads

FAQ

Can I use the MDS02L-05 to provide 5V power for embedded electronics from a 5V-class source?

This converter is specified for a 4.5~5.5Vdc input and provides an unregulated 5V output at 2W. That makes it suitable for low-power embedded 5V electronics when the available source falls within that input range.

Is the MDS02L-05 suitable for 5V industrial interface circuits in compact equipment?

It can be used for 5V interface circuitry in compact equipment where the source is 4.5~5.5Vdc and the load fits within the 2W rating. The 19.5 × 9.8 × 12.5 mm dimensions support use in dense industrial or embedded layouts.

Can this model support low-power 5V automation control electronics?

Yes, it is intended for an unregulated 5V output and a 2W power level from a 4.5~5.5Vdc input. That profile fits low-power automation control electronics rather than higher-power 5V subsystems.

Is the MDS02L-05 meant for outdoor installations?

The published operating temperature range is -40°C to +85°C, so the temperature specification is broad. No IP rating is provided in the specs, so it should be considered for protected indoor or enclosed equipment use rather than direct outdoor exposure.

How does the MDS02L-05 compare with other MDS02L versions?

All MDS02L models are 2W with 4.5~5.5V input. The series includes MDS02L-12 at 12V and 17~167mA and MDS02L-15 at 15V and 13~133mA, while this MDS02L-05 version is 5V and 40~400mA.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Union Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8442608288043, "variants" :[{ "context":"default", "variant_id": 45711600976171, "sku":"MDS02L-05", "inventory_qty" : 0, "customer" :false, "custom_html" :"\u003ca class=\"text-current logged-out-price logged-out\" href=\"https:\/\/wellforces.co.nz\/customer_authentication\/redirect?locale=en\u0026region_country=NZ\"\u003e\n \u003cspan \u003ePlease \u003c\/span\u003e \u003cspan class=\"login-text\"\u003eLogin\u003c\/span\u003e \u003cspan\u003e for Pricing\u003c\/span\u003e\n \u003c\/a\u003e", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS02L-05 DC-DC Converter - 2W - 4.5~5.5V in 5V out 400mA
Mean Well

Mean Well MDS02L-05 DC-DC Converter - 2W - 4.5~5.5V in 5V out 400mA

MDS02L-05 is the 5V output model of this series: rated power 2W, typical efficiency 77%.

MDS02 and MDD02 series are 2W isolated and unregulated module type medical grade DC-DC converter with SIP 7 package. It features international standard pins, a high efficiency up to 8 5 %, wide working temperature range -40~+ 85℃, 6 KVDC or 4 . 2KVAC I/P-O/P hight isolation voltage, short circuit protection, etc. The models account for different input voltage 5V/12V/24V±10%, and various output voltage, 5V/12V/ 15V for single output and ±5V /±9 V/±12V/±15V for dual outputs, which are suitable for medical systems, ultra low leakage current.

Features

  • SIP7 package with international standard pinout
  • Operating temperature range -40 ~ +85℃
  • Medical safety approved (1xMOPP/2xMOOP) according to ANSI/AAMI ES60601-1
  • Low patient leakage current <2μA
  • Protection: Short circuit(3 second max.)
  • 6KVDC or 4.2VAC hight I/O isolation (Reinforced isolation)
  • Low cost
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
View product